Intrauterine, Vacuum‐Induced, Hemorrhage‐Control Device System Contributing to Blood Loss in Fulminant DIC

<b>Background:</b> Postpartum hemorrhage (PPH) complicates 3%-5% of deliveries. Uterine atony is the most common cause of PPH, and there are several interventions that can be used to improve atony, including uterine massage, uterotonics, and intrauterine balloons. The intrauterine vacuum-induced hemorrhage-control device is a suction device that is placed in the lower uterine segment, which creates negative pressure on the myometrium to induce myometrial contraction. <b>Case:</b> The patient was a 37-year-old G3P1011 female with chronic hypertension who presented for induction of labor at term. Spontaneous rupture of membranes occurred several hours later with simultaneous placental abruption leading to DIC. Urgent cesarean section was performed for fetal distress remote from delivery. Massive transfusion protocol was initiated, and an intrauterine vacuum-induced hemorrhage-control device was placed intravaginally in an attempt to control blood loss. However, rather than acting as a tamponade, the device appeared to draw out a great blood loss with its suction. The patient underwent bilateral uterine artery embolization, and the device was replaced with an intrauterine balloon. Patient was transferred to the ICU, and blood products were transfused until the patient was stabilized. <b>Conclusion:</b> While a preliminary study appeared to show the safety and efficacy of the intrauterine vacuum-induced hemorrhage-control device, its use was limited to uterine hemorrhage secondary to uterine atony. It should be avoided in cases of hemorrhage where coagulopathy is suspected.

Authors

Institutions

Publication Details

Journal
Case Reports in Obstetrics and Gynecology
Published
2025-01-01
DOI
https://doi.org/10.1155/crog/3747372
Primary Topic
Maternal and fetal healthcare
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Intrauterine, Vacuum‐Induced, Hemorrhage‐Control Device System Contributing to Blood Loss in Fulminant DIC

Chad Strittmatter, Bruce Rodgers, Alina Zabolotico, Victoria Kucinski
Case Reports in Obstetrics and Gynecology
Maternal and fetal healthcare
article

Intrauterine, Vacuum‐Induced, Hemorrhage‐Control Device System Contributing to Blood Loss in Fulminant DIC

Chad Strittmatter, Bruce Rodgers, Alina Zabolotico, Victoria Kucinski
article en

Abstract

<b>Background:</b> Postpartum hemorrhage (PPH) complicates 3%-5% of deliveries. Uterine atony is the most common cause of PPH, and there are several interventions that can be used to improve atony, including uterine massage, uterotonics, and intrauterine balloons. The intrauterine vacuum-induced hemorrhage-control device is a suction device that is placed in the lower uterine segment, which creates negative pressure on the myometrium to induce myometrial contraction. <b>Case:</b> The patient was a 37-year-old G3P1011 female with chronic hypertension who presented for induction of labor at term. Spontaneous rupture of membranes occurred several hours later with simultaneous placental abruption leading to DIC. Urgent cesarean section was performed for fetal distress remote from delivery. Massive transfusion protocol was initiated, and an intrauterine vacuum-induced hemorrhage-control device was placed intravaginally in an attempt to control blood loss. However, rather than acting as a tamponade, the device appeared to draw out a great blood loss with its suction. The patient underwent bilateral uterine artery embolization, and the device was replaced with an intrauterine balloon. Patient was transferred to the ICU, and blood products were transfused until the patient was stabilized. <b>Conclusion:</b> While a preliminary study appeared to show the safety and efficacy of the intrauterine vacuum-induced hemorrhage-control device, its use was limited to uterine hemorrhage secondary to uterine atony. It should be avoided in cases of hemorrhage where coagulopathy is suspected.

Case Reports in Obstetrics and GynecologyVol. 2025(1)
Sisters of Charity Hospital (US)
Openalex Percentile: Top 67%
Maternal and fetal healthcare
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.