Time-Domain Analysis of Low- and High-Frequency Near-Infrared Spectroscopy Sensor Technologies for Characterization of Cerebral Pressure–Flow and Oxygen Delivery Physiology: A Prospective Observational Study

Cerebrovascular reactivity, cerebral autoregulation (CA), and oxygen delivery can be measured continuously and in a non-invasive fashion using cerebral near-infrared spectroscopy (NIRS). Although the literature is limited surrounding the difference between signals acquired and derived from low (<100 Hz) and high sampling rates (≥100 Hz). As part of a prospective observational study, we preliminarily explored and assessed the difference in the information provided by two NIRS systems using regional cerebral oxygen saturation and cerebral oximetry index signals at low and high sampling rates. The raw data in two frequencies (down-sampled to 1 Hz using the mean and up-sampled to 250 Hz) were decimated to focus on slow-wave vasogenic fluctuations associated with CA. Then, the data were analyzed using various statistical methods such as the absolute signal difference, Pearson correlation, Bland-Altman agreement, Cross-correlation function, optimal time-series autocorrelative structure, time-series impulse response function, and Granger causality relationships. The results of the various statistical analyses indicated that the signals obtained using high-frequency NIRS were different from signals obtained from low-frequency NIRS of the same cerebral region. Hence, high-frequency NIRS systems may possibly contain better signal features compared to NIRS systems with low sampling rates, but further work is required to assess high-frequency NIRS in other healthy and cranial trauma populations.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-09-01
DOI
https://doi.org/10.3390/s25175391
Primary Topic
Optical Imaging and Spectroscopy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Time-Domain Analysis of Low- and High-Frequency Near-Infrared Spectroscopy Sensor Technologies for Characterization of Cerebral Pressure–Flow and Oxygen Delivery Physiology: A Prospective Observational Study

Kevin Y. Stein, Jae Moon, Frederick Adam Zeiler, Amanjyot Singh Sainbhi et al.
Sensors
Optical Imaging and Spectroscopy Techniques
article

Time-Domain Analysis of Low- and High-Frequency Near-Infrared Spectroscopy Sensor Technologies for Characterization of Cerebral Pressure–Flow and Oxygen Delivery Physiology: A Prospective Observational Study

Kevin Y. Stein, Jae Moon, Frederick Adam Zeiler, Amanjyot Singh Sainbhi, Nuray Vakitbilir, Mansoor Hayat, Tobias Bergmann, Rakibul Hasan, Noah Silvaggio
article en

Abstract

Cerebrovascular reactivity, cerebral autoregulation (CA), and oxygen delivery can be measured continuously and in a non-invasive fashion using cerebral near-infrared spectroscopy (NIRS). Although the literature is limited surrounding the difference between signals acquired and derived from low (<100 Hz) and high sampling rates (≥100 Hz). As part of a prospective observational study, we preliminarily explored and assessed the difference in the information provided by two NIRS systems using regional cerebral oxygen saturation and cerebral oximetry index signals at low and high sampling rates. The raw data in two frequencies (down-sampled to 1 Hz using the mean and up-sampled to 250 Hz) were decimated to focus on slow-wave vasogenic fluctuations associated with CA. Then, the data were analyzed using various statistical methods such as the absolute signal difference, Pearson correlation, Bland-Altman agreement, Cross-correlation function, optimal time-series autocorrelative structure, time-series impulse response function, and Granger causality relationships. The results of the various statistical analyses indicated that the signals obtained using high-frequency NIRS were different from signals obtained from low-frequency NIRS of the same cerebral region. Hence, high-frequency NIRS systems may possibly contain better signal features compared to NIRS systems with low sampling rates, but further work is required to assess high-frequency NIRS in other healthy and cranial trauma populations.

SensorsVol. 25(17)
University of Cambridge (GB), Karolinska Institutet (SE), Pan Am Clinic (CA), Addenbrooke's Hospital (GB), University of Manitoba (CA)
Health Sciences Centre Foundation, Research Manitoba, Canadian Institutes of Health Research, Natural Sciences and Engineering Research Council of Canada
Openalex Percentile: Top 67%
Optical Imaging and Spectroscopy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.