ARE THERE BENEfiTS IN TREATING GLIOBLASTOMAS WITH HYPERARC VERSUS RAPIDARC?

Abstract AIMS This study aimed to compare treatment of frontotemporal glioblastoma using standard planar RapidArc versus non-coplanar radiotherapy (HyperArc). METHODS Ten patients with frontotemporal glioblastoma were retrospectively replanned using HyperArc, maintaining original target volumes and dose constraints. Dosimetric parameters, including PTV coverage and doses to or- gans at risk (OARs), were analysed. Isodose contours (10-60 Gy in 10 Gy intervals) were compared. Additional metrics included conformity index (CI), gradient index (GI), planning and delivery time, and overall cost effi- ciency. RESULTS PTV coverage was comparable across both techniques. HyperArc demonstrated superior plan conformity (median CI: 1.06 vs 1.10) and steeper dose fall-off (median GI: 2.41 vs 2.89). HyperArc significantly reduced radiation expo- sure to normal brain tissue, with notable reductions in V20 (27.8%, ~150 cm³), V30 (26.8%, ~75 cm³), V40 (18.5%, ~30 cm³), and V50 (16.3%, ~15 cm³). Mean brain dose was lowered by 8.7% (from 21.5 Gy to 19.8 Gy). Dose to hair-bearing scalp (≥12 Gy) decreased by 25%. However, HyperArc incurred longer delivery times (~10 minutes vs 3–4 minutes), higher energy demands (one plan required 70% more monitor units), and reliance on Encompass shells. CONCLUSION HyperArc improved plan conformity and decreased the volume of normal brain tissue receiving 20-50 Gy. The mean brain dose was reduced to below 20 Gy, which may help mitigate late neurocognitive effects. Dose to scalp associated with alopecia was reduced by approximately 25%, potentially improving cosmetic outcomes. These dosimetric advantages must be weighed against increased treatment complexity, energy use, and operational de- mands.

Authors

Institutions

Publication Details

Journal
Neuro-Oncology
Published
2025-09-01
DOI
https://doi.org/10.1093/neuonc/noaf185.092
Citations
1
Primary Topic
Glioma Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
1.63
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

ARE THERE BENEfiTS IN TREATING GLIOBLASTOMAS WITH HYPERARC VERSUS RAPIDARC?

John Pierre Greenwood, Matthew Ulewich, Louise Dulley, A. Ameen et al.
1 citations
Neuro-Oncology
Glioma Diagnosis and Treatment
1.63
article

ARE THERE BENEfiTS IN TREATING GLIOBLASTOMAS WITH HYPERARC VERSUS RAPIDARC?

John Pierre Greenwood, Matthew Ulewich, Louise Dulley, A. Ameen, Ewan Almond, Debbie Farmakidis
article en
1 citations

Abstract

Abstract AIMS This study aimed to compare treatment of frontotemporal glioblastoma using standard planar RapidArc versus non-coplanar radiotherapy (HyperArc). METHODS Ten patients with frontotemporal glioblastoma were retrospectively replanned using HyperArc, maintaining original target volumes and dose constraints. Dosimetric parameters, including PTV coverage and doses to or- gans at risk (OARs), were analysed. Isodose contours (10-60 Gy in 10 Gy intervals) were compared. Additional metrics included conformity index (CI), gradient index (GI), planning and delivery time, and overall cost effi- ciency. RESULTS PTV coverage was comparable across both techniques. HyperArc demonstrated superior plan conformity (median CI: 1.06 vs 1.10) and steeper dose fall-off (median GI: 2.41 vs 2.89). HyperArc significantly reduced radiation expo- sure to normal brain tissue, with notable reductions in V20 (27.8%, ~150 cm³), V30 (26.8%, ~75 cm³), V40 (18.5%, ~30 cm³), and V50 (16.3%, ~15 cm³). Mean brain dose was lowered by 8.7% (from 21.5 Gy to 19.8 Gy). Dose to hair-bearing scalp (≥12 Gy) decreased by 25%. However, HyperArc incurred longer delivery times (~10 minutes vs 3–4 minutes), higher energy demands (one plan required 70% more monitor units), and reliance on Encompass shells. CONCLUSION HyperArc improved plan conformity and decreased the volume of normal brain tissue receiving 20-50 Gy. The mean brain dose was reduced to below 20 Gy, which may help mitigate late neurocognitive effects. Dose to scalp associated with alopecia was reduced by approximately 25%, potentially improving cosmetic outcomes. These dosimetric advantages must be weighed against increased treatment complexity, energy use, and operational de- mands.

Neuro-OncologyVol. 27(Supplement_2)
Barking, Havering And Redbridge University Hospitals NHS Trust (GB)
Openalex Percentile: Top 14%
Glioma Diagnosis and Treatment
1.63
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.