Feasibility of prototype diamond detectors for pulsed UHDR PBS small-field proton dosimetry for proton FLASH experiments

<i>Objective.</i>This study aims to investigate the responses of prototype diamond detectors under pulsed ultra-high dose rates (UHDRs) pencil-beam-scanning (PBS) protons from a compact proton synchrocyclotron (IBA Proteus®ONE) for small-field UHDR dosimetry.<i>Approach</i>. flashDiamond detectors (fDs) were cross-calibrated with their relative proton responses characterized at conventional dose rates (CONV). Then, absolute UHDR dosimetry was performed and small-field response assessed. These experiments were also conducted with Razor Diode and microdiamond detectors (mDs) for cross-reference. Cross-calibrations were performed against an ADCL-calibrated PPC05 plane-parallel ionization chamber with 59.23 cGy nC<sup>-1</sup>calibration coefficients. fD's linearity, dose-rate, energy, and linear-energy-transfer (LET) responses were assessed under CONV protons. Pulsed UHDR PBS protons of 228 MeV were produced from a medical proton synchrocyclotron (IBA Proteus®ONE) for 1.5 × 1.5-3.0 × 3.0 cm<sup>2</sup>square fields. Nominal absolute UHDR dosimetry was performed at 3 × 3 cm<sup>2</sup>field sizes with relative responses at smaller fields benchmarked against it.<i>Main results</i>. fD had 28.6 ± 0.1 cGy nC<sup>-1</sup>sensitivities under CONV protons and were linear in response with dose-rate independence within ±0.50%. fD were similar to mD in proton energy and LET responses. However, there is an over-response of approximately 5.49%, 6.51% and 13.7% at the 226, 150 and 70 MeV Bragg peaks respectively. Under pulsed proton UHDR irradiation (0.80% s.t.d, 32.6 ± 0.5 cGy dose-per-pulse), fD responded within ±1% as PPC05 with negligible saturation. fD agreed within ±1% with other comparable small-field detectors under small-field UHDR beams and within ±2% of RayStation treatment planning system calculations. There is negligible partial volume averaging with fDs.<i>Significance</i>. Novel fD detectors did not saturate under pulsed UHDR PBS proton irradiation. Their miniscule active crystals make them suitable for small-field dosimetry but render them relatively insensitive compared to mDs. When cross-calibrated, they are suitable for absolute small-field UHDR dosimetry or for relative exit dosimetry monitoring purposes during UHDR radiobiological experiments.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-09-02
DOI
https://doi.org/10.1088/1361-6560/ae023b
Citations
3
Primary Topic
Diamond and Carbon-based Materials Research
Type
article
Field-Weighted Citation Impact
1.55

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Feasibility of prototype diamond detectors for pulsed UHDR PBS small-field proton dosimetry for proton FLASH experiments

Marek Szczepański, Jufri Setianegara, Aoxiang Wang, Hao Gao et al.
3 citations
Physics in Medicine and Biology
Diamond and Carbon-based Materials Research
1.55
article

Feasibility of prototype diamond detectors for pulsed UHDR PBS small-field proton dosimetry for proton FLASH experiments

Marek Szczepański, Jufri Setianegara, Aoxiang Wang, Hao Gao, Yuting Lin, Jarrick Nys, N. Gérard
article en
3 citations

Abstract

<i>Objective.</i>This study aims to investigate the responses of prototype diamond detectors under pulsed ultra-high dose rates (UHDRs) pencil-beam-scanning (PBS) protons from a compact proton synchrocyclotron (IBA Proteus®ONE) for small-field UHDR dosimetry.<i>Approach</i>. flashDiamond detectors (fDs) were cross-calibrated with their relative proton responses characterized at conventional dose rates (CONV). Then, absolute UHDR dosimetry was performed and small-field response assessed. These experiments were also conducted with Razor Diode and microdiamond detectors (mDs) for cross-reference. Cross-calibrations were performed against an ADCL-calibrated PPC05 plane-parallel ionization chamber with 59.23 cGy nC<sup>-1</sup>calibration coefficients. fD's linearity, dose-rate, energy, and linear-energy-transfer (LET) responses were assessed under CONV protons. Pulsed UHDR PBS protons of 228 MeV were produced from a medical proton synchrocyclotron (IBA Proteus®ONE) for 1.5 × 1.5-3.0 × 3.0 cm<sup>2</sup>square fields. Nominal absolute UHDR dosimetry was performed at 3 × 3 cm<sup>2</sup>field sizes with relative responses at smaller fields benchmarked against it.<i>Main results</i>. fD had 28.6 ± 0.1 cGy nC<sup>-1</sup>sensitivities under CONV protons and were linear in response with dose-rate independence within ±0.50%. fD were similar to mD in proton energy and LET responses. However, there is an over-response of approximately 5.49%, 6.51% and 13.7% at the 226, 150 and 70 MeV Bragg peaks respectively. Under pulsed proton UHDR irradiation (0.80% s.t.d, 32.6 ± 0.5 cGy dose-per-pulse), fD responded within ±1% as PPC05 with negligible saturation. fD agreed within ±1% with other comparable small-field detectors under small-field UHDR beams and within ±2% of RayStation treatment planning system calculations. There is negligible partial volume averaging with fDs.<i>Significance</i>. Novel fD detectors did not saturate under pulsed UHDR PBS proton irradiation. Their miniscule active crystals make them suitable for small-field dosimetry but render them relatively insensitive compared to mDs. When cross-calibrated, they are suitable for absolute small-field UHDR dosimetry or for relative exit dosimetry monitoring purposes during UHDR radiobiological experiments.

Physics in Medicine and BiologyVol. 70(19)
Fraunhofer Institute for Physical Measurement Techniques (DE), Ion Beam Applications (Belgium) (BE), University of Kansas Medical Center (US), Huazhong University of Science and Technology (CN), University of Pennsylvania (US), The University of Texas Southwestern Medical Center (US)
National Cancer Institute
Openalex Percentile: Top 16%
Diamond and Carbon-based Materials Research
1.55
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.