Background: Radiation dose generated from computed tomography (CT) has drawn more attention to diagnostic radiology. It is a known fact that the risk of radiation-induced cancer is increasing, thereby necessitating the optimisation of dose in CT protocols. This study focused on determining the radiation dose of CT scans for the abdomen and lung using a 64-slice CT scanner to evaluate their correlation with body mass index (BMI). The objective of the study was to critically evaluate the relationship between BMI and radiation dose metrics in both CT lung and CT abdomen examinations. Methods: Data from 106 patients who underwent CT lung and CT abdomen examinations at an advanced diagnostic center were retrospectively analysed. The volume CT dose index (CTDIvol), dose-length product (DLP), the scan range, and cranium diameter [antero-posterior (AP) and lateral (LAT)] of the patients were documented for further analysis. Effective dose (E) and size-specific dose estimate (SSDE) were also computed. Results: The mean BMI for CT lung was recorded as 24.85 (5.65). However, the correlation between BMI and the dose metrics (SSDE, E, DLP, and CTDIvol) was not significant, with correlation coefficients of 0.1278, 0.047, 0.047, and 0.1147, respectively. In contrast, the BMI for CT abdomen scans showed a moderate correlation with E (0.5898), SSDE (0.6288), DLP (0.5898), and CTDIvol (0.612). The results demonstrate that BMI can be used as a radiation dose metric in the case of CT abdomen scans, but has no influence on CT lung scans. Conclusion: These results further suggest that BMI could provide radiation dose analysis, which in turn leads to optimisation of CT scan parameters.

Authors

Institutions

Publication Details

Journal
Malaysian Journal of Medical Sciences
Published
2025-08-30
DOI
https://doi.org/10.21315/mjms-04-2025-280
Primary Topic
Radiomics and Machine Learning in Medical Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Radiation Dose of Abdominal and Lung Computed Tomography Based on Body Mass Index as an Indicator

Izdihar Kamal, Farouk Kabir Umar, Juliana Mohd Radzi, Muhammad Khalis Abdul Karim et al.
Malaysian Journal of Medical Sciences
Radiomics and Machine Learning in Medical Imaging
article

Radiation Dose of Abdominal and Lung Computed Tomography Based on Body Mass Index as an Indicator

Izdihar Kamal, Farouk Kabir Umar, Juliana Mohd Radzi, Muhammad Khalis Abdul Karim, Mazliana Ahmad Kamarudin, Nor Azura Muhammad, Muhammad Kabeer Sulayman, Bashiru Lukuman, Che Azura Che Abdullah
article en

Abstract

Background: Radiation dose generated from computed tomography (CT) has drawn more attention to diagnostic radiology. It is a known fact that the risk of radiation-induced cancer is increasing, thereby necessitating the optimisation of dose in CT protocols. This study focused on determining the radiation dose of CT scans for the abdomen and lung using a 64-slice CT scanner to evaluate their correlation with body mass index (BMI). The objective of the study was to critically evaluate the relationship between BMI and radiation dose metrics in both CT lung and CT abdomen examinations. Methods: Data from 106 patients who underwent CT lung and CT abdomen examinations at an advanced diagnostic center were retrospectively analysed. The volume CT dose index (CTDIvol), dose-length product (DLP), the scan range, and cranium diameter [antero-posterior (AP) and lateral (LAT)] of the patients were documented for further analysis. Effective dose (E) and size-specific dose estimate (SSDE) were also computed. Results: The mean BMI for CT lung was recorded as 24.85 (5.65). However, the correlation between BMI and the dose metrics (SSDE, E, DLP, and CTDIvol) was not significant, with correlation coefficients of 0.1278, 0.047, 0.047, and 0.1147, respectively. In contrast, the BMI for CT abdomen scans showed a moderate correlation with E (0.5898), SSDE (0.6288), DLP (0.5898), and CTDIvol (0.612). The results demonstrate that BMI can be used as a radiation dose metric in the case of CT abdomen scans, but has no influence on CT lung scans. Conclusion: These results further suggest that BMI could provide radiation dose analysis, which in turn leads to optimisation of CT scan parameters.

Malaysian Journal of Medical SciencesVol. 32(4)
Universiti Putra Malaysia (MY), University of Malaya (MY), Usmanu Danfodiyo University Teaching Hospital, Sokoto (NG), Sultan Zainal Abidin University (MY)
Openalex Percentile: Top 68%
Radiomics and Machine Learning in Medical Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.