Performance Study of the Vibrating Wire Technique to Determine Longitudinal Magnetic Field Profile Using Scans to High Wire Harmonic

Particle accelerator laboratories, which enable world-class research across many scientific fields, depend on the magnets used to manipulate their particle beams for successful operation. The community employs various techniques, typically based on Hall probes and induction sensors/coils, to verify the performance of these accelerator magnets. When the transverse access around a magnet is restricted, conventional Hall probe systems cannot be deployed or require significant modification, while moving wire/coil systems tend to provide information only on the magnetic field’s integral. This research builds upon a vibrating wire setup first commissioned to locate the magnetic center of quadrupole magnets. Scans up to the n = 200 wire harmonic (∼10 kHz drive frequency) were measured to reconstruct the magnetic field across a wire strung through a test magnet. New software was developed to systematically process the many frequency response scans needed for a detailed field reconstruction. This research investigated the speed and precision of the measurement, identifying limitations due to both instrumentation and nonlinear wire behavior. The vibrating wire data agreed with a reference Hall probe scan on the order of 6%; roughly 0.7% RMS error persisted after calibrating the vibrating wire data to the reference scan via scaling factor.

Authors

Institutions

Publication Details

Journal
Metrology
Published
2025-09-01
DOI
https://doi.org/10.3390/metrology5030053
Primary Topic
Non-Destructive Testing Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance Study of the Vibrating Wire Technique to Determine Longitudinal Magnetic Field Profile Using Scans to High Wire Harmonic

Cameron Baribeau
Metrology
Non-Destructive Testing Techniques
article

Performance Study of the Vibrating Wire Technique to Determine Longitudinal Magnetic Field Profile Using Scans to High Wire Harmonic

Cameron Baribeau
article en

Abstract

Particle accelerator laboratories, which enable world-class research across many scientific fields, depend on the magnets used to manipulate their particle beams for successful operation. The community employs various techniques, typically based on Hall probes and induction sensors/coils, to verify the performance of these accelerator magnets. When the transverse access around a magnet is restricted, conventional Hall probe systems cannot be deployed or require significant modification, while moving wire/coil systems tend to provide information only on the magnetic field’s integral. This research builds upon a vibrating wire setup first commissioned to locate the magnetic center of quadrupole magnets. Scans up to the n = 200 wire harmonic (∼10 kHz drive frequency) were measured to reconstruct the magnetic field across a wire strung through a test magnet. New software was developed to systematically process the many frequency response scans needed for a detailed field reconstruction. This research investigated the speed and precision of the measurement, identifying limitations due to both instrumentation and nonlinear wire behavior. The vibrating wire data agreed with a reference Hall probe scan on the order of 6%; roughly 0.7% RMS error persisted after calibrating the vibrating wire data to the reference scan via scaling factor.

MetrologyVol. 5(3)
Canadian Light Source (Canada) (CA)
University of Saskatchewan, Canada Foundation for Innovation, Canadian Institutes of Health Research, Natural Sciences and Engineering Research Council of Canada, National Research Council Canada
Openalex Percentile: Top 77%
Non-Destructive Testing Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.