Scanned particle‐beam tracking with beam correction based on predictive volumetric imaging: A simulation study
The proposed tracking irradiation technique based on predictive volumetric imaging in spot-scanning particle therapy demonstrated tracking doses comparable to doses in the treatment plan across all latency conditions in the lung, liver, and pancreas. Further research and development of treatment devices and treatment planning protocols are warranted for the proposed tracking irradiation technique to become an effective motion management technique in terms of both dosimetric accuracy and treatment efficiency.