Feasibility and limitations of a postal dosimetry audit system using RPLGD for high-dose-rate brachytherapy

International recommendations and South Korean regulations emphasize the need for dosimetry audit systems to ensure treatment accuracy in high-dose-rate brachytherapy. In this study, we designed and tested a prototype polycarbonate postal audit phantom to assess its feasibility. Using six GD-302M radiophotoluminescent glass dosimeters. Dose measurements were performed at points A and B based on the treatment planning system (TPS) and PHITS. These were validated in the Korea laboratory accreditation scheme and secondary standard dosimetry laboratory using Co-60 beam. Reproducibility and linearity tests showed stable responses; however, a maximum deviation of 13.5 % was observed between the measured and TPS doses at point A. The PHITS simulations revealed that the TPS overestimated the dose by approximately 10 %, whereas glass dosimeters showed up to 20 % overestimation because of low-energy gamma rays from the Ir-192 source. Additional water phantom measurements demonstrated that the RPLGD can provide reasonably consistent dose readings under water-equivalent conditions. The discrepancies observed in the audit phantom were found to follow a consistent trend and could be assessed using normalization. The combined expanded uncertainties were calculated as 3.1 % and 3.2 % ( k = 2 ) for the source holder models 70010 and 72280, respectively. The audit phantom showed potential for application in nationwide postal audits, pending further validation. To ensure clinical applicability, future work should focus on addressing the energy dependence of the RPLGD and conducting a multi-institutional pilot study. • Many studies recommend audit systems to verify treatment accuracy through accredited organizations. • Lack of regulatory standards for dosimetry audits in South Korea may lead to safety risks. • A prototype postal dosimetry audit phantom was developed and tested to assess feasibility for future QA systems. • Uncertainty assessment was performed to evaluate the potential applicability of the system for postal audit purposes.

Authors

Institutions

Publication Details

Journal
Radiation Physics and Chemistry
Published
2025-09-01
DOI
https://doi.org/10.1016/j.radphyschem.2025.113285
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Feasibility and limitations of a postal dosimetry audit system using RPLGD for high-dose-rate brachytherapy

Hyeok-Jun Gwon, Sang Hyoun Choi, Kum Bae Kim, Wonho Lee
Radiation Physics and Chemistry
Advanced Radiotherapy Techniques
article

Feasibility and limitations of a postal dosimetry audit system using RPLGD for high-dose-rate brachytherapy

Hyeok-Jun Gwon, Sang Hyoun Choi, Kum Bae Kim, Wonho Lee
article en

Abstract

International recommendations and South Korean regulations emphasize the need for dosimetry audit systems to ensure treatment accuracy in high-dose-rate brachytherapy. In this study, we designed and tested a prototype polycarbonate postal audit phantom to assess its feasibility. Using six GD-302M radiophotoluminescent glass dosimeters. Dose measurements were performed at points A and B based on the treatment planning system (TPS) and PHITS. These were validated in the Korea laboratory accreditation scheme and secondary standard dosimetry laboratory using Co-60 beam. Reproducibility and linearity tests showed stable responses; however, a maximum deviation of 13.5 % was observed between the measured and TPS doses at point A. The PHITS simulations revealed that the TPS overestimated the dose by approximately 10 %, whereas glass dosimeters showed up to 20 % overestimation because of low-energy gamma rays from the Ir-192 source. Additional water phantom measurements demonstrated that the RPLGD can provide reasonably consistent dose readings under water-equivalent conditions. The discrepancies observed in the audit phantom were found to follow a consistent trend and could be assessed using normalization. The combined expanded uncertainties were calculated as 3.1 % and 3.2 % ( k = 2 ) for the source holder models 70010 and 72280, respectively. The audit phantom showed potential for application in nationwide postal audits, pending further validation. To ensure clinical applicability, future work should focus on addressing the energy dependence of the RPLGD and conducting a multi-institutional pilot study. • Many studies recommend audit systems to verify treatment accuracy through accredited organizations. • Lack of regulatory standards for dosimetry audits in South Korea may lead to safety risks. • A prototype postal dosimetry audit phantom was developed and tested to assess feasibility for future QA systems. • Uncertainty assessment was performed to evaluate the potential applicability of the system for postal audit purposes.

Radiation Physics and ChemistryVol. 239
Korea University (KR), Korea Research Institute of Standards and Science (KR), Korea Institute of Radiological and Medical Sciences (KR)
Korea Foundation, Nuclear Safety and Security Commission, Korea Institute of Radiological and Medical Sciences, Ministry of Science and ICT, South Korea
Openalex Percentile: Top 70%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.