Continuous Cuffless Blood Pressure Estimation via Effective and Efficient Broad Learning Model

Hypertension is a critical cardiovascular risk factor, underscoring the necessity of accessible blood pressure (BP) monitoring for its prevention, detection, and management. While cuffless BP estimation using wearable cardiovascular signals via deep learning models (DLMs) offers a promising solution, their implementation often entails high computational costs. This study addresses these challenges by proposing an end-to-end broad learning model (BLM) for efficient cuffless BP estimation. Unlike DLMs that prioritize network depth, the BLM increases network width, thereby reducing computational complexity and enhancing training efficiency for continuous BP estimation. An incremental learning mode is also explored to provide high memory efficiency and flexibility. Validation on the University of California Irvine (UCI) database (403.67 hours) demonstrated that the standard BLM (SBLM) achieved a mean absolute error (MAE) of 11.72 mmHg for arterial BP (ABP) waveform estimation, performance comparable to DLMs such as long short-term memory (LSTM) and the one-dimensional convolutional neural network (1D-CNN), while improving training efficiency by 25.20 times. The incremental BLM (IBLM) offered horizontal scalability by expanding through node addition in a single layer, maintaining predictive performance while reducing storage demands through support for incremental learning with streaming or partial datasets. For systolic and diastolic BP prediction, the SBLM achieved MAEs (mean error $\pm$ standard deviation) of 3.04 mmHg (2.85 $\pm$ 4.15 mmHg) and 2.57 mmHg (-2.47 $\pm$ 3.03 mmHg), respectively. This study highlights the potential of BLM for personalized, real-time, continuous cuffless BP monitoring, presenting a practical solution for healthcare applications.

Authors

Institutions

Publication Details

Journal
IEEE Journal of Biomedical and Health Informatics
Published
2025-09-01
DOI
https://doi.org/10.1109/jbhi.2025.3604464
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Continuous Cuffless Blood Pressure Estimation via Effective and Efficient Broad Learning Model

PingAn Hu, Xiaorong Ding, Zhan Shen, Chunlin Zhang
IEEE Journal of Biomedical and Health Informatics
Non-Invasive Vital Sign Monitoring
article

Continuous Cuffless Blood Pressure Estimation via Effective and Efficient Broad Learning Model

PingAn Hu, Xiaorong Ding, Zhan Shen, Chunlin Zhang
article en

Abstract

Hypertension is a critical cardiovascular risk factor, underscoring the necessity of accessible blood pressure (BP) monitoring for its prevention, detection, and management. While cuffless BP estimation using wearable cardiovascular signals via deep learning models (DLMs) offers a promising solution, their implementation often entails high computational costs. This study addresses these challenges by proposing an end-to-end broad learning model (BLM) for efficient cuffless BP estimation. Unlike DLMs that prioritize network depth, the BLM increases network width, thereby reducing computational complexity and enhancing training efficiency for continuous BP estimation. An incremental learning mode is also explored to provide high memory efficiency and flexibility. Validation on the University of California Irvine (UCI) database (403.67 hours) demonstrated that the standard BLM (SBLM) achieved a mean absolute error (MAE) of 11.72 mmHg for arterial BP (ABP) waveform estimation, performance comparable to DLMs such as long short-term memory (LSTM) and the one-dimensional convolutional neural network (1D-CNN), while improving training efficiency by 25.20 times. The incremental BLM (IBLM) offered horizontal scalability by expanding through node addition in a single layer, maintaining predictive performance while reducing storage demands through support for incremental learning with streaming or partial datasets. For systolic and diastolic BP prediction, the SBLM achieved MAEs (mean error $\pm$ standard deviation) of 3.04 mmHg (2.85 $\pm$ 4.15 mmHg) and 2.57 mmHg (-2.47 $\pm$ 3.03 mmHg), respectively. This study highlights the potential of BLM for personalized, real-time, continuous cuffless BP monitoring, presenting a practical solution for healthcare applications.

IEEE Journal of Biomedical and Health InformaticsVol. 30(2)
University of Electronic Science and Technology of China (CN), Huzhou Normal University (CN)
Openalex Percentile: Top 78%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.