Determination of temporal correction factors for a survey meter detector dedicated to radioprotection in radiology<i></i>

Radiometric surveying is essential for assessing radiation levels in facilities that use ionizing radiation, ensuring that individual exposures remain within the dose limits established by current regulations. Ionization chambers (ICs) are crucial tools in this evaluation, and understanding their operational limitations—particularly the effects of temporal dependence—is key to ensuring accurate measurements. This study aimed to verify the temporal dependence of the Fluke Victoreen IC (349 cm³, model 451B-RYR) by analyzing the ambient dose equivalent rate H*(10) over short exposure intervals in its two operational modes: Integrated (Mode A) and Rate (Mode B). Measurements were conducted using N60, N80, and N100 radiation qualities for narrow X-ray beams defined by ISO 4037-1:2019 and characterized at the Laboratory of Radiological Sciences (LCR). The results revealed temporal dependence for times shorter than 0.8 s for N60 and N80, and shorter than 1.5 s for N100 in both operational modes. Correction factors were calculated to adjust the H*(10) values, reducing uncertainties and enabling the definition of cutoff times for each radiometric quality and chamber mode. Mode A exhibited shorter cutoff times for N60 and N100 (0.3 s and 0.4 s, respectively), whereas Mode B proved more suitable for N80 (0.2 s). The highest combined relative uncertainties before stabilization were observed as follows: for N60, 26% in Mode A and 50% in Mode B (both at 0.1 s); for N80, 31% in Mode A and 17% in Mode B (both at 0.5 s); and for N100, 49% in Mode A (0.2 s) and 15% in Mode B (0.6 s). The study concludes that the application of correction factors and cutoff times improves the precision of H*(10) measurements at short exposure intervals, contributing to more reliable radiometric surveys in radiological protection contexts.

Authors

Institutions

Publication Details

Journal
Brazilian Journal of Radiation Sciences
Published
2025-09-01
DOI
https://doi.org/10.15392/2319-0612.2025.2905
Primary Topic
Graphite, nuclear technology, radiation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Determination of temporal correction factors for a survey meter detector dedicated to radioprotection in radiology

Eduardo Silva, Leonardo De Castro Pacífico, Luís Alexandre Gonçalves Magalhães, Vitoria De Lima Ribeiro Baptista
Brazilian Journal of Radiation Sciences
Graphite, nuclear technology, radiation studies
article

Determination of temporal correction factors for a survey meter detector dedicated to radioprotection in radiology

Eduardo Silva, Leonardo De Castro Pacífico, Luís Alexandre Gonçalves Magalhães, Vitoria De Lima Ribeiro Baptista
article en

Abstract

Radiometric surveying is essential for assessing radiation levels in facilities that use ionizing radiation, ensuring that individual exposures remain within the dose limits established by current regulations. Ionization chambers (ICs) are crucial tools in this evaluation, and understanding their operational limitations—particularly the effects of temporal dependence—is key to ensuring accurate measurements. This study aimed to verify the temporal dependence of the Fluke Victoreen IC (349 cm³, model 451B-RYR) by analyzing the ambient dose equivalent rate H*(10) over short exposure intervals in its two operational modes: Integrated (Mode A) and Rate (Mode B). Measurements were conducted using N60, N80, and N100 radiation qualities for narrow X-ray beams defined by ISO 4037-1:2019 and characterized at the Laboratory of Radiological Sciences (LCR). The results revealed temporal dependence for times shorter than 0.8 s for N60 and N80, and shorter than 1.5 s for N100 in both operational modes. Correction factors were calculated to adjust the H*(10) values, reducing uncertainties and enabling the definition of cutoff times for each radiometric quality and chamber mode. Mode A exhibited shorter cutoff times for N60 and N100 (0.3 s and 0.4 s, respectively), whereas Mode B proved more suitable for N80 (0.2 s). The highest combined relative uncertainties before stabilization were observed as follows: for N60, 26% in Mode A and 50% in Mode B (both at 0.1 s); for N80, 31% in Mode A and 17% in Mode B (both at 0.5 s); and for N100, 49% in Mode A (0.2 s) and 15% in Mode B (0.6 s). The study concludes that the application of correction factors and cutoff times improves the precision of H*(10) measurements at short exposure intervals, contributing to more reliable radiometric surveys in radiological protection contexts.

Brazilian Journal of Radiation SciencesVol. 13(3A (Suppl.))
Universidade Federal do Rio de Janeiro (BR), Universidade do Estado do Rio de Janeiro (BR)
Openalex Percentile: Top 83%
Graphite, nuclear technology, radiation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.