Performance comparison of Mindray EH-2090 Urinalysis analyzers and Sysmex UF-5000
In this study, we used manual microscopy as the gold standard, and compared the analytical and clinical performance of EH-2090 and UF-5000 to evaluate their application characteristics in daily clinical practice. Results show the EH-2090 has comparable analytical performance to Sysmex UF-5000. In clinical performance, the EH-2090 shows better counting accuracy for urine formed elements except bacteria. Besides, the EH-2090 could provide more accurate assessment especially in samples with yeast cells, crystals and dysmorphic red blood cells.
Authors
- Zhicheng Wu (ORCID: https://orcid.org/0000-0003-3498-7520)
- Chen Zhang (ORCID: https://orcid.org/0000-0002-5080-9321)
- Qiaoping Xu
- Juan Chen (ORCID: https://orcid.org/0009-0005-5823-2239)
Institutions
- Zhongda Hospital Southeast University (CN)
- Peking University Shenzhen Hospital (CN)
Publication Details
- Journal
- Scandinavian Journal of Clinical and Laboratory Investigation
- Published
- 2025-08-18
- DOI
- https://doi.org/10.1080/00365513.2025.2550350
- Primary Topic
- Urinary Tract Infections Management
- Type
- article
- Field-Weighted Citation Impact
- 0.00