HyperArc brain SRS target coverage variation due to intra-fractional motion for multiple brain lesions of varying separations and sizes

Abstract Purpose: An evaluation was conducted for HyperArc brain stereotactic radiosurgery (SRS) plans to determine the effect of intra-fractional motion on target coverage and dosimetry to understand the safe limits of zero-PTV-margin HyperArc SRS treatments. Material and methods: The Steve phantom was CT scanned with an Encompass mask and imported into the Eclipse Treatment Planning System, version 16·01. Two high-resolution spherical contours (GTV1 and GTV2) were generated with various sizes and separations. The rotational shift was transformed into translation based on the distance from the tumours to the isocentre and the rotational angle between the tumours. The effect of translational shifts corresponding to each rotational shift on dosimetric performance was evaluated for various tumour sizes and distances to the isocentre. Results: The tumour coverage was compared for different tumour sizes and distances to the isocentre due to intra-fractional motion. For tumour sizes less than 6 mm in diameter, the volume coverage changes significantly for separations greater than 5·5 cm from the isocentre. The effects of intra-fractional motion on dosimetry dominate for separations greater than 5·5 cm. Even though the percent volume coverage changes sharply for extreme shifts, the percent dose coverage stays between 70 and 80% for extreme shifts (1·2° in our case) for all lesion sizes. Conclusions: The target-to-isocentre separation and the size of the lesions are two major factors that contribute to significant dosimetric deviations. When the target-to-isocentre distance is within 3·5 cm, D100% is over 90% coverage. Over 90% target coverage is achievable for zero PTV margin in situations where extreme shifts of 1·2° exist.

Authors

Institutions

Publication Details

Journal
Journal of Radiotherapy in Practice
Published
2025-01-01
DOI
https://doi.org/10.1017/s146039692510023x
Citations
1
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
1.36
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

HyperArc brain SRS target coverage variation due to intra-fractional motion for multiple brain lesions of varying separations and sizes

Aydin Sanli, Ernest Osei, Johnson Darko, Runqing Jiang
1 citations
Journal of Radiotherapy in Practice
Medical Imaging Techniques and Applications
1.36
article

HyperArc brain SRS target coverage variation due to intra-fractional motion for multiple brain lesions of varying separations and sizes

Aydin Sanli, Ernest Osei, Johnson Darko, Runqing Jiang
article en
1 citations

Abstract

Abstract Purpose: An evaluation was conducted for HyperArc brain stereotactic radiosurgery (SRS) plans to determine the effect of intra-fractional motion on target coverage and dosimetry to understand the safe limits of zero-PTV-margin HyperArc SRS treatments. Material and methods: The Steve phantom was CT scanned with an Encompass mask and imported into the Eclipse Treatment Planning System, version 16·01. Two high-resolution spherical contours (GTV1 and GTV2) were generated with various sizes and separations. The rotational shift was transformed into translation based on the distance from the tumours to the isocentre and the rotational angle between the tumours. The effect of translational shifts corresponding to each rotational shift on dosimetric performance was evaluated for various tumour sizes and distances to the isocentre. Results: The tumour coverage was compared for different tumour sizes and distances to the isocentre due to intra-fractional motion. For tumour sizes less than 6 mm in diameter, the volume coverage changes significantly for separations greater than 5·5 cm from the isocentre. The effects of intra-fractional motion on dosimetry dominate for separations greater than 5·5 cm. Even though the percent volume coverage changes sharply for extreme shifts, the percent dose coverage stays between 70 and 80% for extreme shifts (1·2° in our case) for all lesion sizes. Conclusions: The target-to-isocentre separation and the size of the lesions are two major factors that contribute to significant dosimetric deviations. When the target-to-isocentre distance is within 3·5 cm, D100% is over 90% coverage. Over 90% target coverage is achievable for zero PTV margin in situations where extreme shifts of 1·2° exist.

Journal of Radiotherapy in PracticeVol. 24
Grand River Hospital (CA)
Openalex Percentile: Top 16%
Medical Imaging Techniques and Applications
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.