Evaluating the effectiveness of distortion self‐correction for <scp>CEST</scp> ‐ <scp>EPI</scp>
DISC-CEST-EPI can correct the distortion in CEST-EPI at 3 T, leading to improved SSIM and CEST quantification. This approach has the potential to enhance image quality and improve diagnostic accuracy for single-shot CEST-EPI at especially low-field MRI. However, the performance of DISC-CEST-EPI is limited at 11.7 T, and further advancements are necessary to enhance its functionality at ultra-high fields.
Authors
- Yang Liu (ORCID: https://orcid.org/0000-0002-0420-2211)
- Kannie W. Y. Chan (ORCID: https://orcid.org/0000-0002-7315-1550)
- Licheng Ju (ORCID: https://orcid.org/0000-0002-3695-6843)
- Jianpan Huang (ORCID: https://orcid.org/0000-0002-4453-8764)
- Jiadi Xu (ORCID: https://orcid.org/0000-0001-9698-5622)
- Se Weon Park (ORCID: https://orcid.org/0000-0002-4794-0388)
- Lok Hin Law (ORCID: https://orcid.org/0000-0002-7829-370X)
- Kexin Wang (ORCID: https://orcid.org/0000-0003-0124-0631)
Institutions
- Tung Wah College (HK)
- Kennedy Krieger Institute (US)
- Johns Hopkins University (US)
- City University of Hong Kong (HK)
- Chinese University of Hong Kong (HK)
- Tung Wah Hospital (CN)
- City University of Hong Kong, Shenzhen Research Institute (CN)
- University of Hong Kong (HK)
Publication Details
- Journal
- Magnetic Resonance in Medicine
- Published
- 2025-08-29
- DOI
- https://doi.org/10.1002/mrm.70048
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- City University of Hong Kong
- National Natural Science Foundation of China
- Innovation and Technology Commission
- University of Hong Kong
- Innovation and Technology Fund