Optimization of the Non-Local Means Algorithm for Breast Diffusion-Weighted Magnetic Resonance Imaging Using a 3D-Printed Breast-Mimicking Phantom

Diffusion-weighted magnetic resonance (DWMR) images were acquired using a custom-designed, 3D-printed breast-mimicking phantom. The smoothing factor of the non-local means (NLM) algorithm was then optimized for noise reduction. Phantoms were fabricated using polylactic acid, polyethylene terephthalate, and various concentrations of polyvinylpyrrolidone. DWMR images were obtained across b-values ranging from zero to 2000 s/mm<sup>2</sup>. Based on image contrast, the NLM algorithm was applied to the b = 1000 s/mm<sup>2</sup> image, testing smoothing factors from 0.001 to 0.150. The NLM algorithm's performance was quantitatively evaluated using a single DWMR image acquired from this custom phantom. At the optimized smoothing factor, the signal-to-noise ratio (SNR) improved from 96.87 ± 3.42 to 215.81 ± 4.18, and the contrast-to-noise ratio (CNR) from 43.63 ± 2.97 to 131.98 ± 3.56, representing 2.22-fold and 3.02-fold enhancements, respectively. No formal statistical tests were conducted as the analysis was based on a single acquisition. The optimized NLM algorithm also outperformed conventional denoising methods-median, Wiener, and total variation-in both noise suppression and contrast preservation. These findings suggest that the NLM algorithm with optimized parameters is likely to be more effective than existing approaches for enhancing breast DWMR image quality. However, further validation using in vivo patient datasets is essential to confirm its diagnostic utility and clinical generalizability because of the absence of tissue heterogeneity, motion, and physiological noise in the phantom environment.

Authors

Institutions

Publication Details

Journal
Life
Published
2025-08-29
DOI
https://doi.org/10.3390/life15091373
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimization of the Non-Local Means Algorithm for Breast Diffusion-Weighted Magnetic Resonance Imaging Using a 3D-Printed Breast-Mimicking Phantom

Seong-Hyeon Kang, Youngjin Lee, Si‐Hyun Park
Life
Advanced MRI Techniques and Applications
article

Optimization of the Non-Local Means Algorithm for Breast Diffusion-Weighted Magnetic Resonance Imaging Using a 3D-Printed Breast-Mimicking Phantom

Seong-Hyeon Kang, Youngjin Lee, Si‐Hyun Park
article en

Abstract

Diffusion-weighted magnetic resonance (DWMR) images were acquired using a custom-designed, 3D-printed breast-mimicking phantom. The smoothing factor of the non-local means (NLM) algorithm was then optimized for noise reduction. Phantoms were fabricated using polylactic acid, polyethylene terephthalate, and various concentrations of polyvinylpyrrolidone. DWMR images were obtained across b-values ranging from zero to 2000 s/mm<sup>2</sup>. Based on image contrast, the NLM algorithm was applied to the b = 1000 s/mm<sup>2</sup> image, testing smoothing factors from 0.001 to 0.150. The NLM algorithm's performance was quantitatively evaluated using a single DWMR image acquired from this custom phantom. At the optimized smoothing factor, the signal-to-noise ratio (SNR) improved from 96.87 ± 3.42 to 215.81 ± 4.18, and the contrast-to-noise ratio (CNR) from 43.63 ± 2.97 to 131.98 ± 3.56, representing 2.22-fold and 3.02-fold enhancements, respectively. No formal statistical tests were conducted as the analysis was based on a single acquisition. The optimized NLM algorithm also outperformed conventional denoising methods-median, Wiener, and total variation-in both noise suppression and contrast preservation. These findings suggest that the NLM algorithm with optimized parameters is likely to be more effective than existing approaches for enhancing breast DWMR image quality. However, further validation using in vivo patient datasets is essential to confirm its diagnostic utility and clinical generalizability because of the absence of tissue heterogeneity, motion, and physiological noise in the phantom environment.

LifeVol. 15(9)
Gachon University (KR)
National Research Foundation of Korea
Openalex Percentile: Top 68%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.