Evaluating the radiation safety effect of ultraviolet-blocking curtains during portable X-ray examinations in the emergency department: a comparative experimental study

Purpose: This study evaluated the radiation shielding effectiveness of general hospital curtains and ultraviolet (UV)-blocking curtains in emergency departments by measuring the spatial distribution of scatter-radiation dose rates during portable X-ray examinations.Methods: Four shielding conditions were assessed: no curtain, a general hospital curtain, a 99.9% UV-blocking curtain (V2), and a combination of two UV-blocking curtains (V2 & V7). Spatial dose rates were measured at six locations surrounding a simulated patient bed: Feet 0°, Right 0°, Horizontal 45°, Feet 0° & Vertical 30°, Right 0° & Vertical 30°, and Horizontal 45° & Vertical 30°. All measurement points were set 150 cm from the phantom center to simulate the distance between the patients and the surrounding beds.Results: At a distance of 150 cm horizontally from the phantom, the spatial dose rate was lower when a 99.9% UV-blocking curtain was used than when a general hospital curtain or no curtain was used. Among all curtain types, the V2 & V7 99.9% UV-blocking curtains demonstrated the highest shielding efficiency. The lowest spatial dose rate was observed at Feet 0°, where a bed tray was present, while the highest was at Right 0° & Vertical 30°.Conclusion: In beds equipped with the V2 & V7 99.9% UV-blocking curtains and tray, radiation exposure can be minimized by positioning personnel at the foot of the patient. These findings support the use of UV-blocking curtains as a convenient shielding method during portable X-ray procedures in emergency departments. The findings may also guide caregivers and patients to safer positions during examinations.

Authors

Publication Details

Journal
Journal of Korean Biological Nursing Science
Published
2025-08-14
DOI
https://doi.org/10.7586/jkbns.25.027
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluating the radiation safety effect of ultraviolet-blocking curtains during portable X-ray examinations in the emergency department: a comparative experimental study

Kyeonghwan Jeong, Mi-Hyun Han, Nam‐Ju Lee, Dongkyung Jung
Journal of Korean Biological Nursing Science
Radiation Dose and Imaging
article

Evaluating the radiation safety effect of ultraviolet-blocking curtains during portable X-ray examinations in the emergency department: a comparative experimental study

Kyeonghwan Jeong, Mi-Hyun Han, Nam‐Ju Lee, Dongkyung Jung
article en

Abstract

Purpose: This study evaluated the radiation shielding effectiveness of general hospital curtains and ultraviolet (UV)-blocking curtains in emergency departments by measuring the spatial distribution of scatter-radiation dose rates during portable X-ray examinations.Methods: Four shielding conditions were assessed: no curtain, a general hospital curtain, a 99.9% UV-blocking curtain (V2), and a combination of two UV-blocking curtains (V2 & V7). Spatial dose rates were measured at six locations surrounding a simulated patient bed: Feet 0°, Right 0°, Horizontal 45°, Feet 0° & Vertical 30°, Right 0° & Vertical 30°, and Horizontal 45° & Vertical 30°. All measurement points were set 150 cm from the phantom center to simulate the distance between the patients and the surrounding beds.Results: At a distance of 150 cm horizontally from the phantom, the spatial dose rate was lower when a 99.9% UV-blocking curtain was used than when a general hospital curtain or no curtain was used. Among all curtain types, the V2 & V7 99.9% UV-blocking curtains demonstrated the highest shielding efficiency. The lowest spatial dose rate was observed at Feet 0°, where a bed tray was present, while the highest was at Right 0° & Vertical 30°.Conclusion: In beds equipped with the V2 & V7 99.9% UV-blocking curtains and tray, radiation exposure can be minimized by positioning personnel at the foot of the patient. These findings support the use of UV-blocking curtains as a convenient shielding method during portable X-ray procedures in emergency departments. The findings may also guide caregivers and patients to safer positions during examinations.

Journal of Korean Biological Nursing ScienceVol. 27(3)
Catholic University of Daegu
Openalex Percentile: Top 68%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.