Evaluation of cross-calibration based on TRS-398 and TG-51 to modified electron beam calibration methods

Abstract Introduction Accurate electron beam dosimetry is crucial for effective radiotherapy treatment. This study aimed to validate modified electron beam calibration methods through a comprehensive cross-calibration analysis against the established IAEA TRS-398 and AAPM TG-51 protocols. Materials and methods A Varian Trilogy linac with electron beam energy of 6, 9, 12, 15, and 18 MeV was used to perform dosimetric assessments using cylindrical (FC65-G and CC13) and parallel-plate (PPC40) ion chambers. The sequential method was employed for cross-calibration at 18 MeV, with each chamber alternatively serving as the reference and field chambers according to TRS-398 (yielding calibration correction factor N TRS –398 ) and TG-51 (yielding calibration correction factor N TG –51 ) protocols. The ratios of N TRS –398 and N TG –51 compared to the calibration correction factors from country SSDL ( N Co ) it ranged from 0.990 to 1.020. Absorbed doses to water per monitor unit (cGy/MU) were calculated at maximum absorption depths. Results For modified calibration methods, the values of N TRS –398 and N TG –51 yielded absorbed dose values between 0.977-1.005 cGy/MU and 0.980-1.009 cGy/MU, respectively. Dose ratios of the modified methods compared to TRS-398 ranged from 0.982 to 1.010, while ratios compared to TG-51 varied between 0.985 and 1.021. The average absorbed dose to water using N TRS –398 and N TG –51 ranged from 0.984-0.996 cGy/MU and 0.986-0.997 cGy/MU, respectively. The results were also compared with previous studies to demonstrate that the modified calibration methods closely align with the established protocols, with discrepancies within the IAEA’s ±2% tolerance threshold. Conclusions The study highlights the importance of cross-calibration in ensuring the accuracy and reliability of modified electron beam calibration methods. These findings suggest that the modified approaches can serve as effective alternatives to traditional protocols, potentially enhancing dosimetric precision and flexibility in clinical radiotherapy settings.

Authors

Institutions

Publication Details

Journal
Polish Journal of Medical Physics And Engineering
Published
2025-08-28
DOI
https://doi.org/10.2478/pjmpe-2025-0024
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of cross-calibration based on TRS-398 and TG-51 to modified electron beam calibration methods

Supriyanto Ardjo Pawiro, F Y Dimitri, Nelly Nelly, Vladimir S. Ban
Polish Journal of Medical Physics And Engineering
Advanced Radiotherapy Techniques
article

Evaluation of cross-calibration based on TRS-398 and TG-51 to modified electron beam calibration methods

Supriyanto Ardjo Pawiro, F Y Dimitri, Nelly Nelly, Vladimir S. Ban
article en

Abstract

Abstract Introduction Accurate electron beam dosimetry is crucial for effective radiotherapy treatment. This study aimed to validate modified electron beam calibration methods through a comprehensive cross-calibration analysis against the established IAEA TRS-398 and AAPM TG-51 protocols. Materials and methods A Varian Trilogy linac with electron beam energy of 6, 9, 12, 15, and 18 MeV was used to perform dosimetric assessments using cylindrical (FC65-G and CC13) and parallel-plate (PPC40) ion chambers. The sequential method was employed for cross-calibration at 18 MeV, with each chamber alternatively serving as the reference and field chambers according to TRS-398 (yielding calibration correction factor N TRS –398 ) and TG-51 (yielding calibration correction factor N TG –51 ) protocols. The ratios of N TRS –398 and N TG –51 compared to the calibration correction factors from country SSDL ( N Co ) it ranged from 0.990 to 1.020. Absorbed doses to water per monitor unit (cGy/MU) were calculated at maximum absorption depths. Results For modified calibration methods, the values of N TRS –398 and N TG –51 yielded absorbed dose values between 0.977-1.005 cGy/MU and 0.980-1.009 cGy/MU, respectively. Dose ratios of the modified methods compared to TRS-398 ranged from 0.982 to 1.010, while ratios compared to TG-51 varied between 0.985 and 1.021. The average absorbed dose to water using N TRS –398 and N TG –51 ranged from 0.984-0.996 cGy/MU and 0.986-0.997 cGy/MU, respectively. The results were also compared with previous studies to demonstrate that the modified calibration methods closely align with the established protocols, with discrepancies within the IAEA’s ±2% tolerance threshold. Conclusions The study highlights the importance of cross-calibration in ensuring the accuracy and reliability of modified electron beam calibration methods. These findings suggest that the modified approaches can serve as effective alternatives to traditional protocols, potentially enhancing dosimetric precision and flexibility in clinical radiotherapy settings.

Polish Journal of Medical Physics And EngineeringVol. 31(3)
University of Indonesia (ID), Documentation Center of Cambodia (KH), Siloam Eye Hospital (KR)
Universitas Indonesia
Openalex Percentile: Top 71%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.