Abstract Introduction In this article, the effectiveness of two commercial metal artifact reduction algorithms, Siemens iMAR and GE MARS, in computed tomography (CT) imaging is evaluated. Metal artifacts, which arise primarily due to the presence of high atomic number metals in clinical imaging, significantly degrade image quality and impede accurate diagnostics. Material and methods The study compares monoenergetic and dual-energy CT reconstruction algorithms by examining their performance on phantom models, including a Gammex Tissue Characterization Phantom and a custom-made spine stabilization system phantom. Quantitative assessments, such as Hounsfield unit analysis were performed. Results The results show that the iterative reconstruction algorithm (iMAR) from Siemens offers superior artifact suppression and image clarity compared to GE’s dual-energy algorithm (MARS), particularly in cases involving titanium implants. Quantitative assessments, such as Hounsfield unit measurements and visual image analysis, confirm that iMAR produces images with reduced artifacts and more consistent tissue characterization. Conclusions These findings suggest that the choice of artifact reduction algorithm has a profound impact on the diagnostic and planning accuracy of CT scans in patients with metal implants.

Authors

Institutions

Publication Details

Journal
Polish Journal of Medical Physics And Engineering
Published
2025-08-28
DOI
https://doi.org/10.2478/pjmpe-2025-0022
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative Evaluation of Metal Artifact Reduction Algorithms in Computed Tomography: Siemens iMAR vs. GE MARS for Improved Radiotherapy Planning

K. Kisielewicz, Łukasz Brandt, Kamila Nowosad, A. Dziecichowicz et al.
Polish Journal of Medical Physics And Engineering
Advanced X-ray and CT Imaging
article

Comparative Evaluation of Metal Artifact Reduction Algorithms in Computed Tomography: Siemens iMAR vs. GE MARS for Improved Radiotherapy Planning

K. Kisielewicz, Łukasz Brandt, Kamila Nowosad, A. Dziecichowicz, Mansoor M. Ahmed, B. Kiełtyka, Anna Pędracka, K. Rawojć
article en

Abstract

Abstract Introduction In this article, the effectiveness of two commercial metal artifact reduction algorithms, Siemens iMAR and GE MARS, in computed tomography (CT) imaging is evaluated. Metal artifacts, which arise primarily due to the presence of high atomic number metals in clinical imaging, significantly degrade image quality and impede accurate diagnostics. Material and methods The study compares monoenergetic and dual-energy CT reconstruction algorithms by examining their performance on phantom models, including a Gammex Tissue Characterization Phantom and a custom-made spine stabilization system phantom. Quantitative assessments, such as Hounsfield unit analysis were performed. Results The results show that the iterative reconstruction algorithm (iMAR) from Siemens offers superior artifact suppression and image clarity compared to GE’s dual-energy algorithm (MARS), particularly in cases involving titanium implants. Quantitative assessments, such as Hounsfield unit measurements and visual image analysis, confirm that iMAR produces images with reduced artifacts and more consistent tissue characterization. Conclusions These findings suggest that the choice of artifact reduction algorithm has a profound impact on the diagnostic and planning accuracy of CT scans in patients with metal implants.

Polish Journal of Medical Physics And EngineeringVol. 31(3)
Jagiellonian University (PL), Albert Einstein College of Medicine (US), National Institute of Oncology (HU), The Maria Sklodowska-Curie National Research Institute of Oncology (PL), AGH University of Krakow (PL)
Openalex Percentile: Top 79%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.