Comparison of human point-of-care glucometer (Guide Me) and lactometer (Lactate Plus) to an automated chemistry analyzer for measurement of blood concentrations of lactate and glucose in juvenile commercial pigs

Porcine glucose and lactate concentrations measured with a point-of-care glucometer (POC-Glu; Guide Me) and a POC lactometer (POC-Lac; Lactate Plus) were compared to those from a comparative method (CM; COBAS-501c). Fresh whole blood samples (<i>n</i> = 175 glucose; <i>n</i> = 272 lactate) from 10 healthy juvenile commercial-cross pigs (5 barrows, 5 gilts; 79-91-d-old) were collected over 12 d under various conditions. Comparisons were made with the Passing-Bablok regression and Bland-Altman method with multiple measurements per subject. The allowable total error (TE<sub>a</sub>) was set at 20% for glucose and 40% for lactate. Correlation with CM was high for POC-Glu (<i>r</i> = 0.886; 95% CI [0.849, 0.914]) and very high for POC-Lac (<i>r</i> = 0.935; 95% CI [0.918, 0.949]) with constant and proportional biases (intercept = -0.52, 95% CI [-0.98, -0.13] mmol/L, slope = 0.06, 95% CI [0.06, 0.07] mmol/L for glucose; intercept = 0.075, 95% CI [0.050, 0.104] mmol/L, slope = 0.797, 95% CI [0.775, 0.819] mmol/L for lactate). Acceptance limits based on combined inherent imprecision (CIP) were ±5.2% for glucose and ±8.1% for lactate. The observed total errors (TE<sub>obs</sub>) were 5.42% for POC-Glu and 29.9% for POC-Lac at the decision threshold of 3.0 mmol/L, both within the set TE<sub>a</sub>. Both POC-Glu and POC-Lac are practical and reliable tools for use in juvenile pigs, with satisfactory agreement to the reference method, although interpretation of POC-Lac results requires greater caution due to a significant matrix effect, underscoring the need for method- and matrix-specific RIs, particularly for lactate.

Authors

Institutions

Publication Details

Journal
Journal of Veterinary Diagnostic Investigation
Published
2025-08-29
DOI
https://doi.org/10.1177/10406387251366277
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of human point-of-care glucometer (Guide Me) and lactometer (Lactate Plus) to an automated chemistry analyzer for measurement of blood concentrations of lactate and glucose in juvenile commercial pigs

Cary M. Springer, Christopher K. Smith, Gregory A. Shanks, Stephanie A. Kleine et al.
Journal of Veterinary Diagnostic Investigation
Clinical Laboratory Practices and Quality Control
article

Comparison of human point-of-care glucometer (Guide Me) and lactometer (Lactate Plus) to an automated chemistry analyzer for measurement of blood concentrations of lactate and glucose in juvenile commercial pigs

Cary M. Springer, Christopher K. Smith, Gregory A. Shanks, Stephanie A. Kleine, Luca Giori, Bente Flatland, Chiara E. Hampton, Lucille Ruth Vanecek
article en

Abstract

Porcine glucose and lactate concentrations measured with a point-of-care glucometer (POC-Glu; Guide Me) and a POC lactometer (POC-Lac; Lactate Plus) were compared to those from a comparative method (CM; COBAS-501c). Fresh whole blood samples (<i>n</i> = 175 glucose; <i>n</i> = 272 lactate) from 10 healthy juvenile commercial-cross pigs (5 barrows, 5 gilts; 79-91-d-old) were collected over 12 d under various conditions. Comparisons were made with the Passing-Bablok regression and Bland-Altman method with multiple measurements per subject. The allowable total error (TE<sub>a</sub>) was set at 20% for glucose and 40% for lactate. Correlation with CM was high for POC-Glu (<i>r</i> = 0.886; 95% CI [0.849, 0.914]) and very high for POC-Lac (<i>r</i> = 0.935; 95% CI [0.918, 0.949]) with constant and proportional biases (intercept = -0.52, 95% CI [-0.98, -0.13] mmol/L, slope = 0.06, 95% CI [0.06, 0.07] mmol/L for glucose; intercept = 0.075, 95% CI [0.050, 0.104] mmol/L, slope = 0.797, 95% CI [0.775, 0.819] mmol/L for lactate). Acceptance limits based on combined inherent imprecision (CIP) were ±5.2% for glucose and ±8.1% for lactate. The observed total errors (TE<sub>obs</sub>) were 5.42% for POC-Glu and 29.9% for POC-Lac at the decision threshold of 3.0 mmol/L, both within the set TE<sub>a</sub>. Both POC-Glu and POC-Lac are practical and reliable tools for use in juvenile pigs, with satisfactory agreement to the reference method, although interpretation of POC-Lac results requires greater caution due to a significant matrix effect, underscoring the need for method- and matrix-specific RIs, particularly for lactate.

Journal of Veterinary Diagnostic InvestigationVol. 37(6)
Knoxville College (US), University of Tennessee at Knoxville (US)
College of Veterinary Medicine, University of Tennessee
Openalex Percentile: Top 71%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.