Elemental Impurity Detection Technique in Fluorodeoxyglucose (18F) Using Inductively Coupled Plasma Mass Spectrometry: Development and Validation

INTRODUCTION . When developing and validating manufacturing of a product, Russian radiopharmaceutical manufacturers are required to detect ubiquitous impurities (Pb, As, Fe as well as other elements) causing contamination. At the stage of preliminary monitoring, Al, Cu, and Zn were found to most often contribute to fluorodeoxyglucose ( 18 F) elemental contamination. AIM . This study aimed to develop and validate a technique detecting Al, As, Cu, Fe, Pb, and Zn in fluorodeoxyglucose ( 18 F) using inductively coupled plasma mass spectrometry (ICP-MS). MATERIALS AND METHODS . The technique was developed on a model mixture simulating fluorodeoxyglucose (18F). A sample with minimum analysed impurities was used for validation; then varying amounts of standard analyte solutions were added. The content was determined using Agilent 7900 mass spectrometer. Signal intensities of the following isotopes (a.u.m.) were recorded: 27 Al, 57 Fe, 63 Cu, 66 Zn, 75 As, and 208 Pb. RESULTS . The organic matrix was found to insignificantly affect elemental analysis. Due to matrix effect, NaCl reduced spike recovery of Al, Cu, Pb, and Zn to the level below 70%. When diluting model mixture tenfold, spike recovery of the analytes reached 90–101%. Correlation coefficients of element concentration — signal intensity linear dependences for the analytes exceeded 0.99. When estimating precision, RSD values did not exceed 10%. CONCLUSIONS . A technique has been developed detecting Al, As, Cu, Fe, Pb, and Zn in fluorodeoxyglucose ( 18 F) using ICP-MS. This technique considers influence of matrix components on quantification of target elements when preparing samples. Validation parameters of the technique comply with pharmacopoeial requirements.

Authors

Publication Details

Journal
Regulâtornye issledovaniâ i èkspertiza lekarstvennyh sredstv.
Published
2025-08-28
DOI
https://doi.org/10.30895/1991-2919-2025-15-4-444-450
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Elemental Impurity Detection Technique in Fluorodeoxyglucose (18F) Using Inductively Coupled Plasma Mass Spectrometry: Development and Validation

Yu. N. Shvetsova, В. М. Щукин, А. А. Ерина, Н. Е. Кузьмина
Regulâtornye issledovaniâ i èkspertiza lekarstvennyh sredstv.
Medical Imaging Techniques and Applications
article

Elemental Impurity Detection Technique in Fluorodeoxyglucose (18F) Using Inductively Coupled Plasma Mass Spectrometry: Development and Validation

Yu. N. Shvetsova, В. М. Щукин, А. А. Ерина, Н. Е. Кузьмина
article en

Abstract

INTRODUCTION . When developing and validating manufacturing of a product, Russian radiopharmaceutical manufacturers are required to detect ubiquitous impurities (Pb, As, Fe as well as other elements) causing contamination. At the stage of preliminary monitoring, Al, Cu, and Zn were found to most often contribute to fluorodeoxyglucose ( 18 F) elemental contamination. AIM . This study aimed to develop and validate a technique detecting Al, As, Cu, Fe, Pb, and Zn in fluorodeoxyglucose ( 18 F) using inductively coupled plasma mass spectrometry (ICP-MS). MATERIALS AND METHODS . The technique was developed on a model mixture simulating fluorodeoxyglucose (18F). A sample with minimum analysed impurities was used for validation; then varying amounts of standard analyte solutions were added. The content was determined using Agilent 7900 mass spectrometer. Signal intensities of the following isotopes (a.u.m.) were recorded: 27 Al, 57 Fe, 63 Cu, 66 Zn, 75 As, and 208 Pb. RESULTS . The organic matrix was found to insignificantly affect elemental analysis. Due to matrix effect, NaCl reduced spike recovery of Al, Cu, Pb, and Zn to the level below 70%. When diluting model mixture tenfold, spike recovery of the analytes reached 90–101%. Correlation coefficients of element concentration — signal intensity linear dependences for the analytes exceeded 0.99. When estimating precision, RSD values did not exceed 10%. CONCLUSIONS . A technique has been developed detecting Al, As, Cu, Fe, Pb, and Zn in fluorodeoxyglucose ( 18 F) using ICP-MS. This technique considers influence of matrix components on quantification of target elements when preparing samples. Validation parameters of the technique comply with pharmacopoeial requirements.

Regulâtornye issledovaniâ i èkspertiza lekarstvennyh sredstv.Vol. 15(4)
Openalex Percentile: Top 68%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Elemental Impurity Detection Technique in Fluorodeoxyglucose (18F) Using Inductively Coupled Plasma Mass Spectrometry: Development and Validation — Yu. N. Shvetsova, В. М. Щукин, et al. · Regulâtornye issledovaniâ i èkspertiza lekarstvennyh sredstv. (2025) | TGRS Research Map | TGRS