Removal of a third-generation left atrial appendage occlusion device from the pararenal aorta using a novel endovascular technique
Left atrial appendage occluder devices (LAAODs) offer stroke prevention in patients with atrial fibrillation who have bleeding complications or contraindications to oral anticoagulation. We present a case of successful endovascular removal of a third-generation LAAOD embolized to the pararenal aorta. The patient presented asymptomatically after unrelated imaging demonstrated the LAAOD near the lumbar spine. The LAAOD was removed using a laparoscopic biopsy grasper via a percutaneous right femoral approach, thus avoiding a more complicated open procedure.
Authors
- Timothy C. Oskin
- Chase Green (ORCID: https://orcid.org/0009-0002-7628-7127)
- John L. Crawford
- S. Bonthu
Institutions
- University of North Texas (US)
- Texas Christian University (US)
- University of North Texas Health Science Center (US)
- Texas College (US)
- Texas Health Harris Methodist Hospital Fort Worth (US)
Publication Details
- Journal
- Journal of Vascular Surgery Cases and Innovative Techniques
- Published
- 2025-08-28
- DOI
- https://doi.org/10.1016/j.jvscit.2025.101958
- Primary Topic
- Atrial Fibrillation Management and Outcomes
- Type
- article
- Field-Weighted Citation Impact
- 0.00