Impact of Ultrahigh Sensitivity and Continuous Bed Motion on Performance Characteristics of Biograph Vision Quadra PET/CT

Continuous bed motion (CBM) for long-axial-field-of-view PET/CT enables swift total-body examination of patients. However, the sensitivity profile along the axial field of view (AFOV) varies significantly, and its effect on image quality when combined with CBM remains unexplored. This study assesses the effects of ultrahigh sensitivity (UHS) and CBM on recovery coefficients (RCs) and spatial resolution (SR). <b>Methods:</b> Phantom measurements, performed in static bed and CBM acquisition with different bed speeds (3.5, 7, 14, and 30 mm/s), were reconstructed for both high-sensitivity (HS) and UHS modes, and their quality was assessed using the RC and full width at half maximum of the point-spread function. In addition, 7 clinically referred oncologic patients underwent a combined CBM and UHS [<sup>18</sup>F]FDG PET/CT scan. Metrics derived from lesions and healthy tissues were compared across acquisition modes. <b>Results:</b> Mean RC was lower in both UHS (<4.2%) and CBM (<5.9%) than with HS and static acquisitions, respectively. SR in UHS was slightly deteriorated in the central plane of the AFOV (full width at half maximum difference range, 0.06-0.21 mm) but remained comparable toward the edge (full width at half maximum difference range, 0.03-0.07 mm) to SR in HS. In addition, for CBM, SR differences were negligible (full width at half maximum difference, <0.13 mm). SUVs in healthy tissues (<i>r</i> <sup>2</sup> > 0.94) and lesions (<i>r</i> <sup>2</sup> > 0.99) showed excellent correlation. The coefficient of variation was significantly different for the liver. <b>Conclusion:</b> Elaborate protocols using UHS and CBM could be a step toward swift total-body PET care with only a marginal increase in noise. CBM allows a larger part of the body to be examined and helps to mitigate substantial sensitivity differences along the AFOV, albeit at the cost of slightly lower RC.

Authors

Institutions

Publication Details

Journal
Journal of Nuclear Medicine
Published
2025-08-28
DOI
https://doi.org/10.2967/jnumed.125.270078
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of Ultrahigh Sensitivity and Continuous Bed Motion on Performance Characteristics of Biograph Vision Quadra PET/CT

Riemer H. J. A. Slart, Joyce van Sluis, Johannes H. van Snick, Ronald Boellaard et al.
Journal of Nuclear Medicine
Medical Imaging Techniques and Applications
article

Impact of Ultrahigh Sensitivity and Continuous Bed Motion on Performance Characteristics of Biograph Vision Quadra PET/CT

Riemer H. J. A. Slart, Joyce van Sluis, Johannes H. van Snick, Ronald Boellaard, Charalampos Tsoumpas, Andor W.J.M. Glaudemans, Mostafa Roya, Viet Dao, Antoon T. M. Willemsen
article en

Abstract

Continuous bed motion (CBM) for long-axial-field-of-view PET/CT enables swift total-body examination of patients. However, the sensitivity profile along the axial field of view (AFOV) varies significantly, and its effect on image quality when combined with CBM remains unexplored. This study assesses the effects of ultrahigh sensitivity (UHS) and CBM on recovery coefficients (RCs) and spatial resolution (SR). <b>Methods:</b> Phantom measurements, performed in static bed and CBM acquisition with different bed speeds (3.5, 7, 14, and 30 mm/s), were reconstructed for both high-sensitivity (HS) and UHS modes, and their quality was assessed using the RC and full width at half maximum of the point-spread function. In addition, 7 clinically referred oncologic patients underwent a combined CBM and UHS [<sup>18</sup>F]FDG PET/CT scan. Metrics derived from lesions and healthy tissues were compared across acquisition modes. <b>Results:</b> Mean RC was lower in both UHS (<4.2%) and CBM (<5.9%) than with HS and static acquisitions, respectively. SR in UHS was slightly deteriorated in the central plane of the AFOV (full width at half maximum difference range, 0.06-0.21 mm) but remained comparable toward the edge (full width at half maximum difference range, 0.03-0.07 mm) to SR in HS. In addition, for CBM, SR differences were negligible (full width at half maximum difference, <0.13 mm). SUVs in healthy tissues (<i>r</i> <sup>2</sup> > 0.94) and lesions (<i>r</i> <sup>2</sup> > 0.99) showed excellent correlation. The coefficient of variation was significantly different for the liver. <b>Conclusion:</b> Elaborate protocols using UHS and CBM could be a step toward swift total-body PET care with only a marginal increase in noise. CBM allows a larger part of the body to be examined and helps to mitigate substantial sensitivity differences along the AFOV, albeit at the cost of slightly lower RC.

Journal of Nuclear MedicineVol. 66(10)
University of Leeds (GB), University Medical Center Groningen (NL), University of Groningen (NL), Amsterdam Neuroscience (NL), Amsterdam University Medical Centers (NL), University of Twente (NL)
Climate action
Openalex Percentile: Top 68%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.