Impact of Disinfection and Sterilization on 3D-Printing Resin Performance for Surgical Guides in Cardiac Ablation Surgery

Patient-tailored, 3D-printed surgical guides offer significant potential to improve precision and therapeutic efficacy in cardiac ablation surgery. However, reliable post-sterilization material performance presents a critical yet underexplored barrier to clinical adoption. This study investigates how disinfection and sterilization impact the mechanical and thermal properties of photopolymer resins. Specimens from two 3D-printing resins (Bioflex A80 MB™, 3Dresyns; MED625FLX™, Stratasys) were treated with four combinations of disinfection techniques (low-temperature manual cleaning; high-temperature machine washing) and sterilization techniques (H<sub>2</sub>O<sub>2</sub> vs. autoclaving). We assessed post-sterilization properties by mechanical (material integrity, bending tests), thermal (differential scanning calorimetry, thermogravimetric analysis), and viscoelastic (dynamic mechanical analysis) studies. Statistical analysis was performed using one-way ANOVA with Bonferroni post hoc tests (α = 0.05). From this preliminary study, we conclude that MED625FLX maintains integrity and flexibility across all tested disinfection and sterilization methods. Bioflex A80 MB is only suitable for low-temperature disinfection-sterilization, as high-temperature treatments cause surface cracking. Neither resin is appropriate for cryogenic conditions due to the risk of brittleness. Further research into post-sterilization properties is essential to ensure the safety and clinical reliability of these materials in cardiac procedures.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2025-08-28
DOI
https://doi.org/10.3390/bioengineering12090924
Citations
2
Primary Topic
Anatomy and Medical Technology
Type
article
Field-Weighted Citation Impact
1.35

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of Disinfection and Sterilization on 3D-Printing Resin Performance for Surgical Guides in Cardiac Ablation Surgery

Rani Kronenberger, Afsaneh Amirabadi, Carlo de Asmundis, Ivan Eltsov et al.
2 citations
Bioengineering
Anatomy and Medical Technology
1.35
article

Impact of Disinfection and Sterilization on 3D-Printing Resin Performance for Surgical Guides in Cardiac Ablation Surgery

Rani Kronenberger, Afsaneh Amirabadi, Carlo de Asmundis, Ivan Eltsov, Görkem Eylül Kaya, Ramak Hossein Abadi, Luigi Pannone, Jeroen Soete, Mark La Meir, Ali Gharaviri, Kitty Baert, Gian‐Battista Chierchia, Tom Hauffman, Danny Van Hemelrijck, Kalliopi-Artemi Kalteremidou, Giacomo Talevi, Niko Van den Brande, L Uribe, A Paparella, Rawan Kazma
article en
2 citations

Abstract

Patient-tailored, 3D-printed surgical guides offer significant potential to improve precision and therapeutic efficacy in cardiac ablation surgery. However, reliable post-sterilization material performance presents a critical yet underexplored barrier to clinical adoption. This study investigates how disinfection and sterilization impact the mechanical and thermal properties of photopolymer resins. Specimens from two 3D-printing resins (Bioflex A80 MB™, 3Dresyns; MED625FLX™, Stratasys) were treated with four combinations of disinfection techniques (low-temperature manual cleaning; high-temperature machine washing) and sterilization techniques (H<sub>2</sub>O<sub>2</sub> vs. autoclaving). We assessed post-sterilization properties by mechanical (material integrity, bending tests), thermal (differential scanning calorimetry, thermogravimetric analysis), and viscoelastic (dynamic mechanical analysis) studies. Statistical analysis was performed using one-way ANOVA with Bonferroni post hoc tests (α = 0.05). From this preliminary study, we conclude that MED625FLX maintains integrity and flexibility across all tested disinfection and sterilization methods. Bioflex A80 MB is only suitable for low-temperature disinfection-sterilization, as high-temperature treatments cause surface cracking. Neither resin is appropriate for cryogenic conditions due to the risk of brittleness. Further research into post-sterilization properties is essential to ensure the safety and clinical reliability of these materials in cardiac procedures.

BioengineeringVol. 12(9)
Heart Rhythm Society (US), Vrije Universiteit Brussel (BE), Universitair Ziekenhuis Brussel (BE), KU Leuven (BE)
Vrije Universiteit Brussel
Openalex Percentile: Top 18%
Anatomy and Medical Technology
1.35
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.