Evaluation of accuracy, filter performance, and durability among capnography sampling lines: a bench study

The aim of this bench study was to assess performance between currently available sidestream capnography sampling lines claiming compatibility with a Microstream™ enabled monitor system. A series of bench tests were performed on a selection of commercially available sampling lines (oral-nasal cannulas, PetCO2 masks, and procedural bite blocks) to assess PetCO2 accuracy, filter performance and mechanical durability. For each testing procedure, three separate sampling lines were assessed for five repetitions (n = 15 for each product). A 3D printed oral-nasal model connected to a lung simulator was set to two different breathing modes to simulate “normal” and “shallow” breathing. PetCO2 accuracy measurements were completed at both breathing modes to simulate a variety of clinical scenarios, including oxygen delivery. Filter performance and mechanical durability were assessed for cannula sampling lines. When evaluating PetCO2 accuracy across all sampling line types and breathing conditions, error estimates were lowest for matched-paired sampling lines, compared to cross-paired sampling lines. For oral-nasal cannulas, the matched-paired and three cross-paired performed within the system specification for “normal” breathing conditions. Split design oral-nasal cannulas consistently fell outside specifications under conditions of nasal obstruction. Five cross-paired filters clogged prior to expected threshold, with four demonstrating a clog capacity of less than half the expected value. Almost all sampling lines passed mechanical durability testing. Across all testing conditions, cross-paired sampling lines were observed to fall outside system specifications, suggesting that using sampling lines outside the Microstream™ family may disrupt accuracy of readings when using a Microstream™ enabled capnography monitor.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Monitoring and Computing
Published
2025-08-28
DOI
https://doi.org/10.1007/s10877-025-01346-3
Primary Topic
Speech and Audio Processing
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of accuracy, filter performance, and durability among capnography sampling lines: a bench study

Harold Oglesby, J. P. Mulier, Terry L. Jones, Denis Glozman
Journal of Clinical Monitoring and Computing
Speech and Audio Processing
article

Evaluation of accuracy, filter performance, and durability among capnography sampling lines: a bench study

Harold Oglesby, J. P. Mulier, Terry L. Jones, Denis Glozman
article en

Abstract

The aim of this bench study was to assess performance between currently available sidestream capnography sampling lines claiming compatibility with a Microstream™ enabled monitor system. A series of bench tests were performed on a selection of commercially available sampling lines (oral-nasal cannulas, PetCO2 masks, and procedural bite blocks) to assess PetCO2 accuracy, filter performance and mechanical durability. For each testing procedure, three separate sampling lines were assessed for five repetitions (n = 15 for each product). A 3D printed oral-nasal model connected to a lung simulator was set to two different breathing modes to simulate “normal” and “shallow” breathing. PetCO2 accuracy measurements were completed at both breathing modes to simulate a variety of clinical scenarios, including oxygen delivery. Filter performance and mechanical durability were assessed for cannula sampling lines. When evaluating PetCO2 accuracy across all sampling line types and breathing conditions, error estimates were lowest for matched-paired sampling lines, compared to cross-paired sampling lines. For oral-nasal cannulas, the matched-paired and three cross-paired performed within the system specification for “normal” breathing conditions. Split design oral-nasal cannulas consistently fell outside specifications under conditions of nasal obstruction. Five cross-paired filters clogged prior to expected threshold, with four demonstrating a clog capacity of less than half the expected value. Almost all sampling lines passed mechanical durability testing. Across all testing conditions, cross-paired sampling lines were observed to fall outside system specifications, suggesting that using sampling lines outside the Microstream™ family may disrupt accuracy of readings when using a Microstream™ enabled capnography monitor.

Journal of Clinical Monitoring and ComputingVol. 40(1)
Virginia Commonwealth University (US), Ghent University Hospital (BE), Ghent University (BE), St. Joseph’s Candler Hospital (US), Medtronic (Israel) (IL)
Openalex Percentile: Top 73%
Speech and Audio Processing
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.