Protruding Objects in the Membrane Lung Outlet May Increase Thrombogenicity: Fluid Dynamical Insights

Thrombosis in extracorporeal membrane oxygenation (ECMO) circuit components remains a challenge. Besides blood state and surface properties, flow plays a critical role in hemostasis. In this work, we aimed to study the fluid dynamics of a membrane lung (ML) outlet due to its complex design with pins protruding into the blood flow stream (temperature sensor and cap of purge line), with respect to the potential risk of flow-induced coagulation activation. Large eddy simulations were carried out for blood flow of 1 and 4 L/min. Recirculation bubbles and strong vortical structures developed in this geometry. These flow structures were similar to characteristics of flow past bluff bodies, which facilitate entrapment of platelets that may be activated by the elongational shear rates (> 2,000 s-1), observed near the surface of the temperature sensor for the 4 L/min case. Moreover, a thrombus, extracted from an ECMO circuit, was analyzed by scanning electron microscopy. It is concluded that a review of devices used for ECMO with auxiliary objects protruding into the bloodstream is warranted for improvement of design to reduce the risk of blood trauma and coagulation activation.

Authors

Institutions

Publication Details

Journal
ASAIO Journal
Published
2025-08-26
DOI
https://doi.org/10.1097/mat.0000000000002533
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Protruding Objects in the Membrane Lung Outlet May Increase Thrombogenicity: Fluid Dynamical Insights

Monica Emendi, Lisa Prahl Wittberg, Lars Mikael Broman, Frida Nilsson
ASAIO Journal
Mechanical Circulatory Support Devices
article

Protruding Objects in the Membrane Lung Outlet May Increase Thrombogenicity: Fluid Dynamical Insights

Monica Emendi, Lisa Prahl Wittberg, Lars Mikael Broman, Frida Nilsson
article en

Abstract

Thrombosis in extracorporeal membrane oxygenation (ECMO) circuit components remains a challenge. Besides blood state and surface properties, flow plays a critical role in hemostasis. In this work, we aimed to study the fluid dynamics of a membrane lung (ML) outlet due to its complex design with pins protruding into the blood flow stream (temperature sensor and cap of purge line), with respect to the potential risk of flow-induced coagulation activation. Large eddy simulations were carried out for blood flow of 1 and 4 L/min. Recirculation bubbles and strong vortical structures developed in this geometry. These flow structures were similar to characteristics of flow past bluff bodies, which facilitate entrapment of platelets that may be activated by the elongational shear rates (> 2,000 s-1), observed near the surface of the temperature sensor for the 4 L/min case. Moreover, a thrombus, extracted from an ECMO circuit, was analyzed by scanning electron microscopy. It is concluded that a review of devices used for ECMO with auxiliary objects protruding into the bloodstream is warranted for improvement of design to reduce the risk of blood trauma and coagulation activation.

ASAIO Journal
Karolinska University Hospital (SE), Karolinska Institutet (SE), KTH Royal Institute of Technology (SE)
Openalex Percentile: Top 79%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.