Optimization of Block Sequential Regularized Expectation Maximization Reconstruction Parameters for Striatal SPECT Using a Ring-Type Semiconductor SPECT/CT System
This phantom study aimed to optimize the noise-suppression parameter (β) and the edge-preservation parameter (γ) values in the block sequential regularized expectation maximization (BSREM) reconstruction method for <sup>123</sup>I-2-β-carbomethoxy-3β-(4-iodophenyl)-<i>N</i>-(3-fluoropropyl)nortropane striatal SPECT using a ring-type cadmium-zinc-telluride camera. <b>Methods:</b> We used a striatum phantom filled with an <sup>123</sup>I solution, with adjusted concentrations in the left and right caudate and left and right putamen. The phantom was scanned for 120 min using the StarGuide system, and 20-min SPECT data were extracted from the total acquisition. Using the BSREM method, we generated 42 reconstructed images by varying β and γ. We calculated the percent coefficient of variation and percent contrast for each image. Visual evaluations were conducted by a nuclear medicine physician and 2 technologists. Additionally, we calculated the striatal uptake ratio (SUR) and the ratio of the left caudate SUR to putamen SUR in the left striatum (left striatal ratio). <b>Results:</b> The percent coefficient of variation decreased as the β value increased, whereas percent contrast increased as the γ value increased. We conducted a visual evaluation of 15 reconstructed images selected by physical assessment. Images reconstructed with a β value of 0.03 with a γ value of 5 and those with a β value of 0.05 with a γ value of 10 were deemed suitable for diagnostic purposes because of their accurate depiction of tracer accumulation in each striatal region. SUR and left striatal ratio were comparable for these 2 image sets. <b>Conclusion:</b> Evaluation of the optimal conditions for striatal SPECT imaging with BSREM reconstruction revealed that the optimal β value and γ value combinations were a β value of 0.03 and γ value of 5 and a β value of 0.05 and γ value of 10.
Authors
- Akihito Usui (ORCID: https://orcid.org/0000-0003-1889-7496)
- Kentaro Takanami (ORCID: https://orcid.org/0000-0002-0098-7760)
- Hiroyasu Kodama
- Tatsuo Takeuchi (ORCID: https://orcid.org/0009-0008-8786-515X)
- Tomohiro Kaneta (ORCID: https://orcid.org/0000-0003-1146-5092)
- Yoshitaka Tanaka (ORCID: https://orcid.org/0000-0001-8083-845X)
- Hayato Sato (ORCID: https://orcid.org/0000-0003-2599-1229)
- Hayato Odagiri (ORCID: https://orcid.org/0000-0002-9055-305X)
- Nobuo Yamaguchi
Institutions
- Tohoku University (JP)
- Tohoku University Hospital (JP)
Publication Details
- Journal
- Journal of Nuclear Medicine Technology
- Published
- 2025-08-26
- DOI
- https://doi.org/10.2967/jnmt.125.269869
- Primary Topic
- Medical Imaging Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00