Optimization of Block Sequential Regularized Expectation Maximization Reconstruction Parameters for Striatal SPECT Using a Ring-Type Semiconductor SPECT/CT System

This phantom study aimed to optimize the noise-suppression parameter (β) and the edge-preservation parameter (γ) values in the block sequential regularized expectation maximization (BSREM) reconstruction method for <sup>123</sup>I-2-β-carbomethoxy-3β-(4-iodophenyl)-<i>N</i>-(3-fluoropropyl)nortropane striatal SPECT using a ring-type cadmium-zinc-telluride camera. <b>Methods:</b> We used a striatum phantom filled with an <sup>123</sup>I solution, with adjusted concentrations in the left and right caudate and left and right putamen. The phantom was scanned for 120 min using the StarGuide system, and 20-min SPECT data were extracted from the total acquisition. Using the BSREM method, we generated 42 reconstructed images by varying β and γ. We calculated the percent coefficient of variation and percent contrast for each image. Visual evaluations were conducted by a nuclear medicine physician and 2 technologists. Additionally, we calculated the striatal uptake ratio (SUR) and the ratio of the left caudate SUR to putamen SUR in the left striatum (left striatal ratio). <b>Results:</b> The percent coefficient of variation decreased as the β value increased, whereas percent contrast increased as the γ value increased. We conducted a visual evaluation of 15 reconstructed images selected by physical assessment. Images reconstructed with a β value of 0.03 with a γ value of 5 and those with a β value of 0.05 with a γ value of 10 were deemed suitable for diagnostic purposes because of their accurate depiction of tracer accumulation in each striatal region. SUR and left striatal ratio were comparable for these 2 image sets. <b>Conclusion:</b> Evaluation of the optimal conditions for striatal SPECT imaging with BSREM reconstruction revealed that the optimal β value and γ value combinations were a β value of 0.03 and γ value of 5 and a β value of 0.05 and γ value of 10.

Authors

Institutions

Publication Details

Journal
Journal of Nuclear Medicine Technology
Published
2025-08-26
DOI
https://doi.org/10.2967/jnmt.125.269869
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimization of Block Sequential Regularized Expectation Maximization Reconstruction Parameters for Striatal SPECT Using a Ring-Type Semiconductor SPECT/CT System

Akihito Usui, Kentaro Takanami, Hiroyasu Kodama, Tatsuo Takeuchi et al.
Journal of Nuclear Medicine Technology
Medical Imaging Techniques and Applications
article

Optimization of Block Sequential Regularized Expectation Maximization Reconstruction Parameters for Striatal SPECT Using a Ring-Type Semiconductor SPECT/CT System

Akihito Usui, Kentaro Takanami, Hiroyasu Kodama, Tatsuo Takeuchi, Tomohiro Kaneta, Yoshitaka Tanaka, Hayato Sato, Hayato Odagiri, Nobuo Yamaguchi
article en

Abstract

This phantom study aimed to optimize the noise-suppression parameter (β) and the edge-preservation parameter (γ) values in the block sequential regularized expectation maximization (BSREM) reconstruction method for <sup>123</sup>I-2-β-carbomethoxy-3β-(4-iodophenyl)-<i>N</i>-(3-fluoropropyl)nortropane striatal SPECT using a ring-type cadmium-zinc-telluride camera. <b>Methods:</b> We used a striatum phantom filled with an <sup>123</sup>I solution, with adjusted concentrations in the left and right caudate and left and right putamen. The phantom was scanned for 120 min using the StarGuide system, and 20-min SPECT data were extracted from the total acquisition. Using the BSREM method, we generated 42 reconstructed images by varying β and γ. We calculated the percent coefficient of variation and percent contrast for each image. Visual evaluations were conducted by a nuclear medicine physician and 2 technologists. Additionally, we calculated the striatal uptake ratio (SUR) and the ratio of the left caudate SUR to putamen SUR in the left striatum (left striatal ratio). <b>Results:</b> The percent coefficient of variation decreased as the β value increased, whereas percent contrast increased as the γ value increased. We conducted a visual evaluation of 15 reconstructed images selected by physical assessment. Images reconstructed with a β value of 0.03 with a γ value of 5 and those with a β value of 0.05 with a γ value of 10 were deemed suitable for diagnostic purposes because of their accurate depiction of tracer accumulation in each striatal region. SUR and left striatal ratio were comparable for these 2 image sets. <b>Conclusion:</b> Evaluation of the optimal conditions for striatal SPECT imaging with BSREM reconstruction revealed that the optimal β value and γ value combinations were a β value of 0.03 and γ value of 5 and a β value of 0.05 and γ value of 10.

Journal of Nuclear Medicine TechnologyVol. 54(1)
Tohoku University (JP), Tohoku University Hospital (JP)
Openalex Percentile: Top 69%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.