Creatinine-on-a-chip: colorimetric ELISA-based serum creatinine detection in a microfluidic device

Chronic kidney diseases (CKDs), which often end in kidney failure for many people around the world, have an important place in public health given that they also trigger other diseases. Therefore, the development of fast and cost-effective diagnostic technologies enables effective monitoring of patients and early diagnosis. Here, using the Enzyme-Linked Immunosorbent Assay (ELISA) principle, serum creatinine concentrations were determined using the developed lab-on-a-chip (LOC) platform. In this system, which was termed "<i>creatinine-on-a-chip"</i>, colorimetric ELISA protocol was applied to determine creatinine levels in a microfluidic chip functionalized with creatinine-specific antibodies. Creatinine detection was performed by quantifying the absorbance difference between the detection and reference channels, normalized to the reference signal within the microfluidic chip. The detection signal intensity varied depending on the region selected along the microfluidic channel. The adsorption of the capture antibody used for surface functionalization, which was particularly more pronounced near the inlet region, played a critical role in the detection signal. These findings suggest that random selection of the detection area can lead to significant signal variability, and that careful selection of a well-characterized region is essential for improving detection performance. With this developed system, creatinine was detected with high sensitivity in the linear range of 1-20 μg mL<sup>-1</sup>, both spiked in phosphate buffered saline (PBS) and fetal bovine serum (FBS). Using the creatinine-on-a-chip, serum creatinine analysis can be performed rapidly (∼15 min) in a cost-effective manner ($1.05 per test).

Authors

Institutions

Publication Details

Journal
The Analyst
Published
2025-01-01
DOI
https://doi.org/10.1039/d5an00650c
Citations
1
Primary Topic
Electrochemical sensors and biosensors
Type
article
Field-Weighted Citation Impact
0.73
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Creatinine-on-a-chip: colorimetric ELISA-based serum creatinine detection in a microfluidic device

Betul Karakuzu, H. Cumhur Tekin
1 citations
The Analyst
Electrochemical sensors and biosensors
0.73
article

Creatinine-on-a-chip: colorimetric ELISA-based serum creatinine detection in a microfluidic device

Betul Karakuzu, H. Cumhur Tekin
article en
1 citations

Abstract

Chronic kidney diseases (CKDs), which often end in kidney failure for many people around the world, have an important place in public health given that they also trigger other diseases. Therefore, the development of fast and cost-effective diagnostic technologies enables effective monitoring of patients and early diagnosis. Here, using the Enzyme-Linked Immunosorbent Assay (ELISA) principle, serum creatinine concentrations were determined using the developed lab-on-a-chip (LOC) platform. In this system, which was termed "<i>creatinine-on-a-chip"</i>, colorimetric ELISA protocol was applied to determine creatinine levels in a microfluidic chip functionalized with creatinine-specific antibodies. Creatinine detection was performed by quantifying the absorbance difference between the detection and reference channels, normalized to the reference signal within the microfluidic chip. The detection signal intensity varied depending on the region selected along the microfluidic channel. The adsorption of the capture antibody used for surface functionalization, which was particularly more pronounced near the inlet region, played a critical role in the detection signal. These findings suggest that random selection of the detection area can lead to significant signal variability, and that careful selection of a well-characterized region is essential for improving detection performance. With this developed system, creatinine was detected with high sensitivity in the linear range of 1-20 μg mL<sup>-1</sup>, both spiked in phosphate buffered saline (PBS) and fetal bovine serum (FBS). Using the creatinine-on-a-chip, serum creatinine analysis can be performed rapidly (∼15 min) in a cost-effective manner ($1.05 per test).

The AnalystVol. 150(19)
Izmir Institute of Technology (TR), Middle East Technical University (TR)
Openalex Percentile: Top 24%
Electrochemical sensors and biosensors
0.73
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Creatinine-on-a-chip: colorimetric ELISA-based serum creatinine detection in a microfluidic device — Betul Karakuzu, H. Cumhur Tekin · The Analyst (2025) | TGRS Research Map | TGRS