Comparing fluid dynamics of newborn and adult centrifugal pumps in cardiopulmonary bypass procedures

The newborn pump features notably shorter exposure times (45% lower than the adult design), reduced peak stress values, and a 20% reduction in the volume of fluid exposed to stress levels exceeding 50 Pa, suggesting a potential decrease in the risk of blood damage. Additionally, its reduced extent of stagnation zones (0.13 ml compared to 0.21 ml) indicates improved washout performance, thus lowering the risk of platelet aggregation and thrombus formation. These findings suggest that using a paediatric pump instead of an adult pump at typical flow rates for newborn patients may reduce the risk of blood damage.

Authors

Institutions

Publication Details

Journal
Medical Engineering & Physics
Published
2025-08-25
DOI
https://doi.org/10.1016/j.medengphy.2025.104422
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparing fluid dynamics of newborn and adult centrifugal pumps in cardiopulmonary bypass procedures

Filippo Consolo, Silvia Colombo, Paolo Fontanili, Alberto Redaelli et al.
Medical Engineering & Physics
Mechanical Circulatory Support Devices
article

Comparing fluid dynamics of newborn and adult centrifugal pumps in cardiopulmonary bypass procedures

Filippo Consolo, Silvia Colombo, Paolo Fontanili, Alberto Redaelli, Silvia Bozzi, Marco Corbelli, Giuseppe Passoni, Federica Bisconti
article en

Abstract

The newborn pump features notably shorter exposure times (45% lower than the adult design), reduced peak stress values, and a 20% reduction in the volume of fluid exposed to stress levels exceeding 50 Pa, suggesting a potential decrease in the risk of blood damage. Additionally, its reduced extent of stagnation zones (0.13 ml compared to 0.21 ml) indicates improved washout performance, thus lowering the risk of platelet aggregation and thrombus formation. These findings suggest that using a paediatric pump instead of an adult pump at typical flow rates for newborn patients may reduce the risk of blood damage.

Medical Engineering & PhysicsVol. 145(1)
Vita-Salute San Raffaele University (IT), Eurosets (Italy) (IT), Politecnico di Milano (IT)
Politecnico di Milano
Openalex Percentile: Top 79%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Comparing fluid dynamics of newborn and adult centrifugal pumps in cardiopulmonary bypass procedures — Filippo Consolo, Silvia Colombo, et al. · Medical Engineering & Physics (2025) | TGRS Research Map | TGRS