Hb Tacoma is difficult to detect and may cause false results with different HbA1c methods

The haemoglobin variant Hb Tacoma [HBB:c.93G > T;CD30(Arg > Ser) was detected in approximately 1 in 1000 patients tested for HbA1c at Karolinska University Hospital, Stockholm. Incidentally, five homozygous individuals were found, none of them presenting with severe haematological abnormalities. When the cation exchange chromatographic method Bio-Rad Variant II was used, chromatograms representing Hb Tacoma heterozygotes were easily misinterpreted as normal, although the HbA1c results were about 40% lower than those obtained by immunoassays. This is explained by the fact that the glycated form of Hb Tacoma coelutes with the LA1c fraction, while the main fraction of Hb Tacoma is not separated from normal HbA. Automatic calculation of the ratio LA1c/HbA1c was soon introduced and samples, where this ratio exceeded certain limits, were further investigated. In case of Hb Tacoma, HbA1c results were obtained from immunoassays. Several other HbA1c methods were evaluated using samples from Hb Tacoma heterozygotes. Unlike the Variant II method, the HbA1c results from the Bio-Rad D-100 method did not significantly differ from immunoassay. However, the results for Hb Tacoma samples deviated in an unpredictable way. The cation exchange chromatographic method Tosoh G11 constantly showed a positive bias. Hb Tacoma did not significantly affect the HbA1c results using immunoassays like Tina-quant Gen. 3 (Roche) and DCA Vantage (Siemens), an affinity method (Afinion 2, Abbott) or an enzymatic method (Alinity c, Abbott). Although the capillary electrophoresis HbA1c method (Sebia Capillarys 3 Tera) did not demonstrate any significant bias, the imprecision for this method was unacceptably high for samples with Hb Tacoma.

Authors

Institutions

Publication Details

Journal
Scandinavian Journal of Clinical and Laboratory Investigation
Published
2025-08-18
DOI
https://doi.org/10.1080/00365513.2025.2547317
Primary Topic
Cardiovascular Function and Risk Factors
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Hb Tacoma is difficult to detect and may cause false results with different HbA1c methods

Niklas Bark, Per Bjellerup, Sara Karlsson, Britta Landin et al.
Scandinavian Journal of Clinical and Laboratory Investigation
Cardiovascular Function and Risk Factors
article

Hb Tacoma is difficult to detect and may cause false results with different HbA1c methods

Niklas Bark, Per Bjellerup, Sara Karlsson, Britta Landin, Mikael Nilsson, Soheir Beshara, Johan Eskils
article en

Abstract

The haemoglobin variant Hb Tacoma [HBB:c.93G > T;CD30(Arg > Ser) was detected in approximately 1 in 1000 patients tested for HbA1c at Karolinska University Hospital, Stockholm. Incidentally, five homozygous individuals were found, none of them presenting with severe haematological abnormalities. When the cation exchange chromatographic method Bio-Rad Variant II was used, chromatograms representing Hb Tacoma heterozygotes were easily misinterpreted as normal, although the HbA1c results were about 40% lower than those obtained by immunoassays. This is explained by the fact that the glycated form of Hb Tacoma coelutes with the LA1c fraction, while the main fraction of Hb Tacoma is not separated from normal HbA. Automatic calculation of the ratio LA1c/HbA1c was soon introduced and samples, where this ratio exceeded certain limits, were further investigated. In case of Hb Tacoma, HbA1c results were obtained from immunoassays. Several other HbA1c methods were evaluated using samples from Hb Tacoma heterozygotes. Unlike the Variant II method, the HbA1c results from the Bio-Rad D-100 method did not significantly differ from immunoassay. However, the results for Hb Tacoma samples deviated in an unpredictable way. The cation exchange chromatographic method Tosoh G11 constantly showed a positive bias. Hb Tacoma did not significantly affect the HbA1c results using immunoassays like Tina-quant Gen. 3 (Roche) and DCA Vantage (Siemens), an affinity method (Afinion 2, Abbott) or an enzymatic method (Alinity c, Abbott). Although the capillary electrophoresis HbA1c method (Sebia Capillarys 3 Tera) did not demonstrate any significant bias, the imprecision for this method was unacceptably high for samples with Hb Tacoma.

Scandinavian Journal of Clinical and Laboratory InvestigationVol. 85(6)
Uppsala University (SE), Karolinska University Hospital (SE), Karolinska Institutet (SE), Region Blekinge (SE), Center for Clinical Research Dalarna (SE), Region Västmanland (SE)
Openalex Percentile: Top 65%
Cardiovascular Function and Risk Factors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.