Low-Power Flexible Classifier Chip for Atrial Fibrillation Detection
This paper outlines a fully integrated on-chip classifier on a flexible substrate for real-time detection of atrial fibrillation (AFib) from electrocardiogram (ECG) signals. The ECG signals are digitized by means of a 14-bit analog-to-digital converter (ADC), followed by a time-domain feature extractor. These features are sent to a switched-capacitor compute-in-memory, analog, three-layer artificial neural network (ANN) for classification. The test-chip, fabricated in 65nm technology, consumes 58.3<i>µ</i>J/ per inference and achieves an average accuracy greater than 99.5% on Physionet dataset, 80% on a small, prospective human study performed at Mayo Clinic, Arizona and 99.9% on a large, retrospective dataset provided by Mayo Clinic Enterprise referred to as Rochester dataset.
Authors
- Arindam Sanyal (ORCID: https://orcid.org/0000-0003-4045-6291)
- Imon Banerjee (ORCID: https://orcid.org/0000-0002-3327-8004)
- Sudarsan Sadasivuni (ORCID: https://orcid.org/0000-0002-9008-8426)
- Jose Sanchez
- Sumukh Prashant Bhanushali (ORCID: https://orcid.org/0000-0001-7463-2949)
Institutions
- Mayo Clinic Hospital (US)
- Arizona State University (US)
- University at Buffalo, State University of New York (US)
Publication Details
- Journal
- IEEE transactions on circuits and systems for artificial intelligence.
- Published
- 2025-08-14
- DOI
- https://doi.org/10.1109/tcasai.2025.3597591
- Primary Topic
- Conducting polymers and applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00