Development of a Fitting Algorithm and Mobile Application for Portable Electrochemical Impedance Spectroscopy

Introduction: Electrochemical impedance spectroscopy (EIS) is a technique that utilizes alternating-current (AC) signals for analyzing charge transfer and chemical reaction kinetics. This technique has several clinical applications including biomolecular detection. While many potentiostats have been developed to make EIS more accessible, such as HELPStat, current portable potentiostat designs with EIS capabilities rely on external hardware for control of the device or for data analysis, limiting designs to non-portable applications. Methods: To address this issue, a Levenberg-Marquardt fitting algorithm was implemented in the HELPStat firmware, allowing approximation of charge transfer resistance from empirical impedance measurements. The developed algorithm was compared to a MATLAB script, and both were evaluated using impedance data for ideal Randles circuits as well as potassium ferricyanide and cortisol aptamer binding Redox reactions. Results were analyzed using paired t-tests comparing absolute value of residuals and root-mean-squared-error (RMSE). In addition, an Android application was developed to perform bidirectional communication between the HELPStat potentiostat and a mobile device using Bluetooth Low-Energy (BLE), with a serial monitor being used to confirm that potentiostat measurements were being configured as expected. Results: Testing of the fitting algorithm demonstrated a statistically significant difference between the developed algorithm and a similar MATLAB implementation. Further analysis revealed that the developed fitting algorithm had statistically comparable or lower errors. In addition, testing with a cortisol aptamer was able to successfully differentiate between 1.5 ng/ml and 7 ng/ml samples of cortisol. By developing a novel Android application and enabling true wireless functionality through BLE, we enable a better bridge of communication between user and device. Conclusion: We developed a fitting algorithm in device firmware along with a novel Android application that allows device control and data analysis with the HELPStat potentiostat. By implementing this without dependence on external hardware, we enable device portability for point-of-care applications.

Authors

Publication Details

Journal
Proceedings of IMPRS
Published
2025-06-24
DOI
https://doi.org/10.18060/29092
Primary Topic
Engineering Applied Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a Fitting Algorithm and Mobile Application for Portable Electrochemical Impedance Spectroscopy

Jacqueline C. Linnes, Kevin Alessandro Bautista, Shannon Riegle
3 links
Proceedings of IMPRS
Engineering Applied Research
article

Development of a Fitting Algorithm and Mobile Application for Portable Electrochemical Impedance Spectroscopy

Jacqueline C. Linnes, Kevin Alessandro Bautista, Shannon Riegle
article en

Abstract

Introduction: Electrochemical impedance spectroscopy (EIS) is a technique that utilizes alternating-current (AC) signals for analyzing charge transfer and chemical reaction kinetics. This technique has several clinical applications including biomolecular detection. While many potentiostats have been developed to make EIS more accessible, such as HELPStat, current portable potentiostat designs with EIS capabilities rely on external hardware for control of the device or for data analysis, limiting designs to non-portable applications. Methods: To address this issue, a Levenberg-Marquardt fitting algorithm was implemented in the HELPStat firmware, allowing approximation of charge transfer resistance from empirical impedance measurements. The developed algorithm was compared to a MATLAB script, and both were evaluated using impedance data for ideal Randles circuits as well as potassium ferricyanide and cortisol aptamer binding Redox reactions. Results were analyzed using paired t-tests comparing absolute value of residuals and root-mean-squared-error (RMSE). In addition, an Android application was developed to perform bidirectional communication between the HELPStat potentiostat and a mobile device using Bluetooth Low-Energy (BLE), with a serial monitor being used to confirm that potentiostat measurements were being configured as expected. Results: Testing of the fitting algorithm demonstrated a statistically significant difference between the developed algorithm and a similar MATLAB implementation. Further analysis revealed that the developed fitting algorithm had statistically comparable or lower errors. In addition, testing with a cortisol aptamer was able to successfully differentiate between 1.5 ng/ml and 7 ng/ml samples of cortisol. By developing a novel Android application and enabling true wireless functionality through BLE, we enable a better bridge of communication between user and device. Conclusion: We developed a fitting algorithm in device firmware along with a novel Android application that allows device control and data analysis with the HELPStat potentiostat. By implementing this without dependence on external hardware, we enable device portability for point-of-care applications.

Proceedings of IMPRSVol. 7(1)
Openalex Percentile: Top 77%
Engineering Applied Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Development of a Fitting Algorithm and Mobile Application for Portable Electrochemical Impedance Spectroscopy — Jacqueline C. Linnes, Kevin Alessandro Bautista, et al. · Proceedings of IMPRS (2025) | TGRS Research Map | TGRS