Counting-loss correction procedure of X-ray imaging detectors with consideration for the effective atomic number of biological objects

It is necessary to correct counting loss caused by the pulse pile-up effect and dead time when using energy-resolving photon-counting detectors (ERPCDs) under “high-counting-rate” conditions in medical and/or industrial settings. We aimed to develop a novel counting-loss correction procedure in which biological objects having effective atomic numbers ( Z eff values) of 6.5–13.0 are measured with polychromatic X-rays. To correct for counting loss, such a procedure must theoretically estimate the count value of an ideal X-ray spectrum without counting loss. In this study, we estimated the ideal X-ray spectrum by focusing on the following two points: (1) the X-ray attenuation in an object ( Z eff values of 6.5–13.0) and (2) the detector response. Virtual materials having intermediate atomic numbers between 6.5 and 13.0 were generated by using a mixture of polymethylmethacrylate (PMMA, Z eff = 6.5) and aluminum (Al, Z eff = 13.0). We then constructed an algorithm that can perform the counting-loss correction based on the object's true Z eff value. To demonstrate the applicability of our procedure, we analyzed investigational objects consisting of PMMA and Al using a prototype ERPCD system. A fresh fish sample was also analyzed. The Z eff values agree with the theoretical values within an accuracy of Z eff ±1. In conclusion, we have developed a highly accurate procedure for correcting counting losses for the quantitative X-ray imaging of biological objects. • A procedure for correcting counting loss in photon-counting detectors was proposed. • The effective atomic number ( Z eff ) of the object was considered for correction. • Our method is applicable to biological objects having Z eff of 6.5–13.0. • Applicability was evaluated experimentally by creating an Z eff image. • The accuracy of the Z eff image was achieved to within ± 1.

Authors

Institutions

Publication Details

Journal
Radiation Physics and Chemistry
Published
2025-08-14
DOI
https://doi.org/10.1016/j.radphyschem.2025.113237
Citations
3
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
2.02

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Counting-loss correction procedure of X-ray imaging detectors with consideration for the effective atomic number of biological objects

Hiroaki Hayashi, Yuki Kanazawa, Natsumi Kimoto, Akitoshi Katsumata et al.
3 citations
Radiation Physics and Chemistry
Advanced X-ray and CT Imaging
2.02
article

Counting-loss correction procedure of X-ray imaging detectors with consideration for the effective atomic number of biological objects

Hiroaki Hayashi, Yuki Kanazawa, Natsumi Kimoto, Akitoshi Katsumata, Rina Nishigami, Daiki Kobayashi, Takashi Asahara, Sota Goto, Shuichiro Yamamoto, Tatsuya Maeda
article en
3 citations

Abstract

It is necessary to correct counting loss caused by the pulse pile-up effect and dead time when using energy-resolving photon-counting detectors (ERPCDs) under “high-counting-rate” conditions in medical and/or industrial settings. We aimed to develop a novel counting-loss correction procedure in which biological objects having effective atomic numbers ( Z eff values) of 6.5–13.0 are measured with polychromatic X-rays. To correct for counting loss, such a procedure must theoretically estimate the count value of an ideal X-ray spectrum without counting loss. In this study, we estimated the ideal X-ray spectrum by focusing on the following two points: (1) the X-ray attenuation in an object ( Z eff values of 6.5–13.0) and (2) the detector response. Virtual materials having intermediate atomic numbers between 6.5 and 13.0 were generated by using a mixture of polymethylmethacrylate (PMMA, Z eff = 6.5) and aluminum (Al, Z eff = 13.0). We then constructed an algorithm that can perform the counting-loss correction based on the object's true Z eff value. To demonstrate the applicability of our procedure, we analyzed investigational objects consisting of PMMA and Al using a prototype ERPCD system. A fresh fish sample was also analyzed. The Z eff values agree with the theoretical values within an accuracy of Z eff ±1. In conclusion, we have developed a highly accurate procedure for correcting counting losses for the quantitative X-ray imaging of biological objects. • A procedure for correcting counting loss in photon-counting detectors was proposed. • The effective atomic number ( Z eff ) of the object was considered for correction. • Our method is applicable to biological objects having Z eff of 6.5–13.0. • Applicability was evaluated experimentally by creating an Z eff image. • The accuracy of the Z eff image was achieved to within ± 1.

Radiation Physics and ChemistryVol. 239
Kanazawa University (JP), Asahi University (JP), Okayama University (JP), Kanazawa Medical University (JP), Kumamoto Health Science University (JP), Kobe Tokiwa University (JP), Junshin Gakuen University (JP), Okayama University Hospital (JP), IHI Corporation (Japan) (JP), Kumamoto University (JP)
Fukuoka Public Health Promotion Organization Cancer Research Fund, Japan Society for the Promotion of Science
Openalex Percentile: Top 13%
Advanced X-ray and CT Imaging
2.02
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.