Method comparison of diabetes mellitus associated autoantibodies in serum specimens
Our study reveals significant and clinically relevant discrepancies in the detection of GADA and IA-2A when comparing CLIA devices to the ELISA method routinely used in our laboratory. For individual antibodies up to 65 % of patients who tested positive using ELISA showed negative or borderline results on at least one CLIA device. These deviations are unpredictable and cannot be detected through standard calibration and internal control procedures.
Authors
- Anna E. Rapp (ORCID: https://orcid.org/0000-0001-5122-765X)
- Stefanie Busch
- Peter Findeisen
- Alexander Lenard
Institutions
- Dr Kolbach Kliniek (NL)
Publication Details
- Journal
- Clinical Chemistry and Laboratory Medicine (CCLM)
- Published
- 2025-08-22
- DOI
- https://doi.org/10.1515/cclm-2025-0137
- Primary Topic
- Diabetes and associated disorders
- Type
- article
- Field-Weighted Citation Impact
- 0.00