Verification of the T50 Calciprotein Crystallization test: bias estimation and interferences
The T50 test was successfully implemented in a routine laboratory setting. Additionally, the precision and interference observed in this study largely agreed with the manufacturer's claims.
Authors
- Andreas Pasch (ORCID: https://orcid.org/0000-0002-7439-0748)
- Vincent Linder
- Marc Vervloet (ORCID: https://orcid.org/0000-0003-4607-9415)
- Róbert de Jonge (ORCID: https://orcid.org/0000-0002-3454-0094)
- Henrike M. Hamer (ORCID: https://orcid.org/0000-0002-6544-2331)
- Joyce Y. Xu (ORCID: https://orcid.org/0009-0000-6257-8797)
- Didier Falconnet
- Tarah van den Berkmortel
- Schelder Gabriel Bertoncello Rosa
Institutions
- Johannes Kepler University of Linz (AT)
- Union Bank of Switzerland (CH)
- Radboud University Nijmegen (NL)
- Radboud University Medical Center (NL)
- Stiftung Lindenhof Bern (CH)
- Amsterdam Neuroscience (NL)
- Radboud Institute for Molecular Life Sciences (NL)
- Lindenhofspital (CH)
- University of Amsterdam (NL)
Publication Details
- Journal
- Clinical Chemistry and Laboratory Medicine (CCLM)
- Published
- 2025-08-20
- DOI
- https://doi.org/10.1515/cclm-2025-0600
- Citations
- 1
- Primary Topic
- Blood properties and coagulation
- Type
- article
- Field-Weighted Citation Impact
- 1.52
Funders
- Eurostars