Impact of Volume Control Biosensors on Blood Pressure During Haemodialysis: A Quasi-Experimental Study

<b>Background:</b> Intradialytic hypotension is a common complication in haemodialysis, affecting up to 30% of sessions. It results from an imbalance between ultrafiltration and compensatory mechanisms, such as vascular tone and plasma refilling. Volume-controlled biosensors allow for the continuous monitoring of the haemoconcentration, enabling early detection and prevention of hypotension. <b>Methods:</b> A quasi-experimental study was conducted to assess the effectiveness of biosensors in reducing hypotensive episodes. Two biosensors were compared: the Blood Volume Monitor and the Haemomaster system. Data were collected over two four-month periods: before and after biosensor implementation. Nursing staff received specific training, and a protocol for consistent data collection was established. Informed consent was obtained from all eligible participants. The incidence of intradialytic hypotension was compared between sessions with and without biosensor use. Additionally, outcomes were analysed according to biosensor type. <b>Results:</b> A total of 2262 dialysis sessions from 22 patients were analysed. The cohort was 54.5% male, with a mean age of 60 years (SE = 21); 27.3% had diabetes and 81.8% had hypertension. Post-dilution haemodiafiltration was performed in 62.8% of sessions. Intradialysis hypotension occurred in 11.2% of sessions using biosensors compared to 14.0% without (<i>p</i> = 0.021). No significant difference was found between biosensor types (10.8% vs. 11.8%; <i>p</i> = 0.531), although device 1 reached a significantly lower critical blood volume (mean: 10 L; SE = 4 vs. 16 L; SE = 5; <i>p</i> = 0.000). <b>Conclusions:</b> Biosensor use was associated with fewer hypotensive episodes and greater haemodynamic stability. These findings support their integration into routine dialysis practice to improve treatment, safety, and individualised care.

Authors

Institutions

Publication Details

Journal
Healthcare
Published
2025-08-20
DOI
https://doi.org/10.3390/healthcare13162060
Primary Topic
Dialysis and Renal Disease Management
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of Volume Control Biosensors on Blood Pressure During Haemodialysis: A Quasi-Experimental Study

Carla Pérez-Ingidua, Ana Ascaso, Ismael Ortuño‐Soriano, A. Paterna et al.
Healthcare
Dialysis and Renal Disease Management
article

Impact of Volume Control Biosensors on Blood Pressure During Haemodialysis: A Quasi-Experimental Study

Carla Pérez-Ingidua, Ana Ascaso, Ismael Ortuño‐Soriano, A. Paterna, Veronica Gimeno-Hernan, Natividad Codesal-Sanabria, Guillermo Perez-Duque
article en

Abstract

<b>Background:</b> Intradialytic hypotension is a common complication in haemodialysis, affecting up to 30% of sessions. It results from an imbalance between ultrafiltration and compensatory mechanisms, such as vascular tone and plasma refilling. Volume-controlled biosensors allow for the continuous monitoring of the haemoconcentration, enabling early detection and prevention of hypotension. <b>Methods:</b> A quasi-experimental study was conducted to assess the effectiveness of biosensors in reducing hypotensive episodes. Two biosensors were compared: the Blood Volume Monitor and the Haemomaster system. Data were collected over two four-month periods: before and after biosensor implementation. Nursing staff received specific training, and a protocol for consistent data collection was established. Informed consent was obtained from all eligible participants. The incidence of intradialytic hypotension was compared between sessions with and without biosensor use. Additionally, outcomes were analysed according to biosensor type. <b>Results:</b> A total of 2262 dialysis sessions from 22 patients were analysed. The cohort was 54.5% male, with a mean age of 60 years (SE = 21); 27.3% had diabetes and 81.8% had hypertension. Post-dilution haemodiafiltration was performed in 62.8% of sessions. Intradialysis hypotension occurred in 11.2% of sessions using biosensors compared to 14.0% without (<i>p</i> = 0.021). No significant difference was found between biosensor types (10.8% vs. 11.8%; <i>p</i> = 0.531), although device 1 reached a significantly lower critical blood volume (mean: 10 L; SE = 4 vs. 16 L; SE = 5; <i>p</i> = 0.000). <b>Conclusions:</b> Biosensor use was associated with fewer hypotensive episodes and greater haemodynamic stability. These findings support their integration into routine dialysis practice to improve treatment, safety, and individualised care.

HealthcareVol. 13(16)
Universidad Complutense de Madrid (ES), Hospital Clínico San Carlos (ES), Instituto de Investigación Sanitaria del Hospital Clínico San Carlos (ES), Hospital Universitario del Henares (ES)
Openalex Percentile: Top 77%
Dialysis and Renal Disease Management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.