Dexamethasone-eluting cochlear implants reduce inflammation and foreign body response in human and murine cochleae

The inflammatory foreign body response (FBR) following cochlear implantation (CI) can negatively impact CI outcomes, including increased electrode impedances. This study aims to investigate the long-term efficacy of dexamethasone-eluting cochlear implant and locally delivered dexamethasone, a potent anti-inflammatory glucocorticoid, on the intracochlear FBR and electrical impedance post-implantation in a murine model. Preliminary impedance data in humans are also provided as a complement to the murine data to illustrate generalizability and reinforce implications related to clinical application. The left ears of CX3CR1<sup>+/GFP</sup> Thy1<sup>+/YFP</sup> (macrophage-neuron dual reporter) mice were implanted with dexamethasone-eluting cochlear implants (Dex-CI) or standard implant (Standard-CI) while the right ear served as unoperated control. Another group of dual reporter mice was implanted with a standard CI electrode array followed by injection of dexamethasone in the middle ear to mimic current clinical practice (Dex-local). Mouse implants were electrically stimulated with serial measurements of electrical impedance. Human subjects were implanted with either standard or Dex-CI followed by serial impedance measurements. Dex-CI reduced electrical impedance in the murine model and human subjects and inflammatory FBR in the murine model for an extended period. Dex-local in the murine model is ineffective for long-term reduction of FBR and electrode impedance. Our data suggests that dexamethasone-eluting arrays are more effective than the current clinical practice of locally applied dexamethasone in reducing FBR and electrical impedance.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-08-20
DOI
https://doi.org/10.1038/s41598-025-10739-y
Citations
5
Primary Topic
Hearing Loss and Rehabilitation
Type
article
Field-Weighted Citation Impact
7.16

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Dexamethasone-eluting cochlear implants reduce inflammation and foreign body response in human and murine cochleae

Robert J. Morell, Alexander D. Claussen, Peter Eckard, Rafal T. Olszewski et al.
5 citations
Scientific Reports
Hearing Loss and Rehabilitation
7.16
article

Dexamethasone-eluting cochlear implants reduce inflammation and foreign body response in human and murine cochleae

Robert J. Morell, Alexander D. Claussen, Peter Eckard, Rafal T. Olszewski, Ya Lang Enke, Shoujun Gu, Muhammad Taifur Rahman, Jacob Oleson, Rachel A. Scheperle, Marlan R. Hansen, Keiko Hirose, Bryce Hunger, Robert D. Gay, Douglas M. Bennion, Ibrahim Razu, Jonathon R. Kirk, Brian Mostaert, Cristina García–Cardona, Michael Hoa, Shakila Mahmuda Fatima, Nashwaan Ali Khan
article en
5 citations

Abstract

The inflammatory foreign body response (FBR) following cochlear implantation (CI) can negatively impact CI outcomes, including increased electrode impedances. This study aims to investigate the long-term efficacy of dexamethasone-eluting cochlear implant and locally delivered dexamethasone, a potent anti-inflammatory glucocorticoid, on the intracochlear FBR and electrical impedance post-implantation in a murine model. Preliminary impedance data in humans are also provided as a complement to the murine data to illustrate generalizability and reinforce implications related to clinical application. The left ears of CX3CR1<sup>+/GFP</sup> Thy1<sup>+/YFP</sup> (macrophage-neuron dual reporter) mice were implanted with dexamethasone-eluting cochlear implants (Dex-CI) or standard implant (Standard-CI) while the right ear served as unoperated control. Another group of dual reporter mice was implanted with a standard CI electrode array followed by injection of dexamethasone in the middle ear to mimic current clinical practice (Dex-local). Mouse implants were electrically stimulated with serial measurements of electrical impedance. Human subjects were implanted with either standard or Dex-CI followed by serial impedance measurements. Dex-CI reduced electrical impedance in the murine model and human subjects and inflammatory FBR in the murine model for an extended period. Dex-local in the murine model is ineffective for long-term reduction of FBR and electrode impedance. Our data suggests that dexamethasone-eluting arrays are more effective than the current clinical practice of locally applied dexamethasone in reducing FBR and electrical impedance.

Scientific ReportsVol. 15(1)
Cochlear (Australia) (AU), University of Iowa (US), Washington University in St. Louis (US), National Institute on Deafness and Other Communication Disorders (US)
Cochlear, National Institutes of Health, National Institute on Deafness and Other Communication Disorders, National Center for Advancing Translational Sciences
Zero hunger
Openalex Percentile: Top 3%
Hearing Loss and Rehabilitation
7.16
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.