Performance of neutron beam system for accelerator-based boron neutron capture therapy

A neutron beam system (NBS) based on an electrostatic proton accelerator designed for accelerator-based boron neutron capture therapy has been installed in Xiamen Humanity Hospital in China. The NBS has been used for patient treatments since October 2023, after more than a year of preliminary studies. By the end of 2024, more than 50 treatments had been executed. NBS uptime during 2024 was above 96%. Repeatability of the beam parameters was well within 2%. At present, configuration with only one treatment room operational, the overall machine usage, including clinical use, physics and dosimetry measurements, and biological experiments, has reached 27% of its capacity.

Authors

Publication Details

Journal
Advances in Radiotherapy & Nuclear Medicine
Published
2025-08-18
DOI
https://doi.org/10.36922/arnm025190022
Citations
1
Primary Topic
Boron Compounds in Chemistry
Type
article
Field-Weighted Citation Impact
1.36
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance of neutron beam system for accelerator-based boron neutron capture therapy

А. Н. Макаров, A. B. Muchnikov, A. Dunaevsky, Alexander A. Ivanov et al.
1 citations
Advances in Radiotherapy & Nuclear Medicine
Boron Compounds in Chemistry
1.36
article

Performance of neutron beam system for accelerator-based boron neutron capture therapy

А. Н. Макаров, A. B. Muchnikov, A. Dunaevsky, Alexander A. Ivanov, K. Y. Franzen, Michael Meekins, Kirill Martianov, Chad Lee, Y. Jiang, Mark Harrison, Vladislav Veksleman
article en
1 citations

Abstract

A neutron beam system (NBS) based on an electrostatic proton accelerator designed for accelerator-based boron neutron capture therapy has been installed in Xiamen Humanity Hospital in China. The NBS has been used for patient treatments since October 2023, after more than a year of preliminary studies. By the end of 2024, more than 50 treatments had been executed. NBS uptime during 2024 was above 96%. Repeatability of the beam parameters was well within 2%. At present, configuration with only one treatment room operational, the overall machine usage, including clinical use, physics and dosimetry measurements, and biological experiments, has reached 27% of its capacity.

Advances in Radiotherapy & Nuclear MedicineVol. 3(4)
Openalex Percentile: Top 17%
Boron Compounds in Chemistry
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.