Integrated Microfluidic Tissue Barrier Sensor Module for a Standardized and Modular Organ-On-Chip Platform

We present the fluidic and electrical packaging of a novel silicon-based trans-epithelial electrical resistance (TEER) sensor chip designed for a modular and standardized organ-on-chip (OoC) platform. The package comprises three key components: the housing of the TEER chip, microfluidic routing for seamless integration with the platform, and electrical connections to a platform-integrated potentiostat. This modular solution enables continuous impedance measurements while maintaining unobstructed optical access to the tissue culture region. Experiments confirmed leak-free fluid flow across the stacked microfluidic channels and stable sensitivity of TiN electrodes to PBS. The TEER module retains optical transparency, bi-ocompatibility, and industrial scalability, supporting advanced in situ tissue barrier assessments in standardized OoC systems.

Authors

Institutions

Publication Details

Published
2025-06-29
DOI
https://doi.org/10.1109/transducers61432.2025.11110020
Primary Topic
3D Printing in Biomedical Research
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Integrated Microfluidic Tissue Barrier Sensor Module for a Standardized and Modular Organ-On-Chip Platform

Mathieu Odijk, Aniruddha Paul, Hans Bouwmeester, Jia-Jun Yeh et al.
3D Printing in Biomedical Research
article

Integrated Microfluidic Tissue Barrier Sensor Module for a Standardized and Modular Organ-On-Chip Platform

Mathieu Odijk, Aniruddha Paul, Hans Bouwmeester, Jia-Jun Yeh, Jaap M. J. den Toonder, Germaine Aalderink, Hande Aydoğmuş, Pratik Tawade, Massimo Mastrangeli
article en

Abstract

We present the fluidic and electrical packaging of a novel silicon-based trans-epithelial electrical resistance (TEER) sensor chip designed for a modular and standardized organ-on-chip (OoC) platform. The package comprises three key components: the housing of the TEER chip, microfluidic routing for seamless integration with the platform, and electrical connections to a platform-integrated potentiostat. This modular solution enables continuous impedance measurements while maintaining unobstructed optical access to the tissue culture region. Experiments confirmed leak-free fluid flow across the stacked microfluidic channels and stable sensitivity of TiN electrodes to PBS. The TEER module retains optical transparency, bi-ocompatibility, and industrial scalability, supporting advanced in situ tissue barrier assessments in standardized OoC systems.

Eindhoven University of Technology (NL), Wageningen University & Research (NL), University of Twente (NL), Delft University of Technology (NL)
Technische Universiteit Delft, Nederlandse Organisatie voor Wetenschappelijk Onderzoek
Openalex Percentile: Top 80%
3D Printing in Biomedical Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.