AI/ML Based Design Optimization & Fabrication of ALN Piezo-Cantilever Array for Bio-Mimetic MEMS Cochlea
We propose a biomimetic, fully implantable MEMS-based cochlear implant utilizing an AlN piezo-cantilever array. Gaussian Process Regression (GPR) integrated with Genetic Algorithm was used to optimize cantilever geometry for maximum charge output at a desired frequency within 0.25 mm<sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> area. Multiple machine learning models were trained with COMSOL data, and GPR gave best prediction with <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">$\mathrm{R}^{2}=0.9999$</tex> for resonant frequency and <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">$\mathrm{R}^{2}=0.98$</tex> for charge. Optimized designs generate 0.7-8 mV, sufficient for direct auditory neuron stimulation. Fabrication employed a new two-step method to grow thick, high quality AlN layer. We also report initial measurement of resonant frequencies and sensitivity of the fabricated array.
Authors
- Pavitra Jain
- Naveen D. Souza N.
- Anju Sebastian (ORCID: https://orcid.org/0000-0001-7517-9904)
- Saurabh A. Chandorkar (ORCID: https://orcid.org/0000-0002-3125-7007)
Institutions
- Indian Institute of Science Bangalore (IN)
Publication Details
- Published
- 2025-06-29
- DOI
- https://doi.org/10.1109/transducers61432.2025.11109211
- Primary Topic
- Innovative Energy Harvesting Technologies
- Type
- article
- Field-Weighted Citation Impact
- 0.00