Ultra-High Frequency and Small Aperture Size Circular Array Based on Piezoelectric Polymer for Intravascular Ultrasound Imaging
This work presents a novel approach for fabricating circular arrays for intravascular ultrasound (IVUS) imaging, successfully achieving a 32-element, 55 MHz transducer with a compact 1.5 mm aperture size, based on piezoelectric polymer. The innovative use of PVDF-TrFE (polyvinylidene fluoride-trifluoroethylene) as the piezoelectric material enables high bandwidth (80%) and minimal crosstalk (< −60 dB). Radial imaging tests were conducted using needles (D = 0.8 mm). The synthetic aperture method produced an image with a 2 mm penetration depth and an estimated a radial −6 dB beamwidth of 30 μm. The circular array provides a promising solution to the challenges of system complexity in high-frequency IVUS imaging.
Authors
- Tao Peng
- Huahuang Luo (ORCID: https://orcid.org/0000-0002-0044-4643)
- Jingwen Yang (ORCID: https://orcid.org/0000-0001-9282-0521)
- Zihan Lu
- Wei-Ning Lee (ORCID: https://orcid.org/0000-0001-8799-2492)
- Kanjie Du (ORCID: https://orcid.org/0009-0005-1166-3533)
- Zhiqing Zhang (ORCID: https://orcid.org/0000-0003-0120-2598)
- Weiting Liu (ORCID: https://orcid.org/0000-0003-0035-2500)
- Qingqing Ke (ORCID: https://orcid.org/0000-0001-9434-1161)
- Chunlong Cheng (ORCID: https://orcid.org/0000-0001-6202-7434)
- Guoxiang Zhang (ORCID: https://orcid.org/0000-0002-4531-2628)
Institutions
- Sun Yat-sen University (CN)
- University of Hong Kong (HK)
Publication Details
- Published
- 2025-06-29
- DOI
- https://doi.org/10.1109/transducers61432.2025.11111396
- Citations
- 1
- Primary Topic
- Advanced Sensor and Energy Harvesting Materials
- Type
- article
- Field-Weighted Citation Impact
- 0.67