Accurate Subcutaneous Glucose Prediction Based on Reverse Iontophoresis with Skin Surface PH Calibration

Reverse iontophoresis (RI)-based approaches face challenges in achieving accurate epidermal-to-subcutaneous biomarkers prediction. Previously, we found that the skin surface pH undergoes significant change during RI, affecting the skin's zeta potential and, consequently, influencing the glucose extraction flux; and the changed pH also affect the performance of the glucose sensor. In this work, we firstly experimentally explore the impact of the change of skin surface pH on glucose extraction by RI and on the glucose sensor performance, thus then making accurate subcutaneous glucose prediction based on pH calibration. A flexible electronic device was developed, incorporating symmetric extraction electrodes for interstitial fluid (ISF) collection and sensors for glucose, Na<sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">+</sup> and pH measurement. The results showed that the device achieves a mean absolute percentage error (MAPE) of 2.9% in glucose prediction with pH calibration, while for 11.2% without calibration. This finding offers key insights into the RI mechanism of ISF extraction and related precise health monitoring.

Authors

Institutions

Publication Details

Published
2025-06-29
DOI
https://doi.org/10.1109/transducers61432.2025.11109521
Primary Topic
Spectroscopy Techniques in Biomedical and Chemical Research
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accurate Subcutaneous Glucose Prediction Based on Reverse Iontophoresis with Skin Surface PH Calibration

Wangwang Zhu, Zhihua Pu, Xi Li, Wen‐Jun Li et al.
Spectroscopy Techniques in Biomedical and Chemical Research
article

Accurate Subcutaneous Glucose Prediction Based on Reverse Iontophoresis with Skin Surface PH Calibration

Wangwang Zhu, Zhihua Pu, Xi Li, Wen‐Jun Li, Xingguo Zhang, Hao Zheng, Chenxi Jin, Dachao Li
article en

Abstract

Reverse iontophoresis (RI)-based approaches face challenges in achieving accurate epidermal-to-subcutaneous biomarkers prediction. Previously, we found that the skin surface pH undergoes significant change during RI, affecting the skin's zeta potential and, consequently, influencing the glucose extraction flux; and the changed pH also affect the performance of the glucose sensor. In this work, we firstly experimentally explore the impact of the change of skin surface pH on glucose extraction by RI and on the glucose sensor performance, thus then making accurate subcutaneous glucose prediction based on pH calibration. A flexible electronic device was developed, incorporating symmetric extraction electrodes for interstitial fluid (ISF) collection and sensors for glucose, Na<sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">+</sup> and pH measurement. The results showed that the device achieves a mean absolute percentage error (MAPE) of 2.9% in glucose prediction with pH calibration, while for 11.2% without calibration. This finding offers key insights into the RI mechanism of ISF extraction and related precise health monitoring.

Tianjin University (CN)
National Natural Science Foundation of China, China Postdoctoral Science Foundation, National Key Research and Development Program of China
Zero hunger
Openalex Percentile: Top 85%
Spectroscopy Techniques in Biomedical and Chemical Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.