Palm-Sized Wireless MEMS Flow Sensor System Attachable to Medical Mask for Real-Time Respiration Monitoring

Pulmonary diseases are of the leading causes of deaths worldwide. This has motivated researchers to develop respiration monitoring devices to allow enhanced prevention and early diagnosis of the diseases. In this study, we propose a novel palm-sized real-time wireless respiration monitoring system attachable to conventional medical masks. A micro-electro mechanical systems (MEMS) calorimetric flow sensor is at the heart of the system. The sensor system is powered with coin batteries and the measurement data was transmitted through Bluetooth. The sensor package was optimized to achieve a stable output even with a short flow channel length of 34 mm, limiting the dead space. A measurement of respiration rate and tidal volume was demonstrated, which was consistent with a spirometry reference generally used for respiration measurements in hospitals.

Authors

Institutions

Publication Details

Published
2025-06-29
DOI
https://doi.org/10.1109/transducers61432.2025.11109997
Primary Topic
Energy Efficient Wireless Sensor Networks
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Palm-Sized Wireless MEMS Flow Sensor System Attachable to Medical Mask for Real-Time Respiration Monitoring

Miyoko Matsushima, Tsuyoshi Tsukada, Yoshihiro Hasegawa, Mitsuhiro Shikida et al.
4 links
Energy Efficient Wireless Sensor Networks
article

Palm-Sized Wireless MEMS Flow Sensor System Attachable to Medical Mask for Real-Time Respiration Monitoring

Miyoko Matsushima, Tsuyoshi Tsukada, Yoshihiro Hasegawa, Mitsuhiro Shikida, Muhammad Salman Al Farisi, Tsutomu Kawabe, Shin Hasegawa
article en

Abstract

Pulmonary diseases are of the leading causes of deaths worldwide. This has motivated researchers to develop respiration monitoring devices to allow enhanced prevention and early diagnosis of the diseases. In this study, we propose a novel palm-sized real-time wireless respiration monitoring system attachable to conventional medical masks. A micro-electro mechanical systems (MEMS) calorimetric flow sensor is at the heart of the system. The sensor system is powered with coin batteries and the measurement data was transmitted through Bluetooth. The sensor package was optimized to achieve a stable output even with a short flow channel length of 34 mm, limiting the dead space. A measurement of respiration rate and tidal volume was demonstrated, which was consistent with a spirometry reference generally used for respiration measurements in hospitals.

Hiroshima City University (JP), Nagoya University (JP), Cosmo Oil (Japan) (JP)
Openalex Percentile: Top 76%
Energy Efficient Wireless Sensor Networks
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Palm-Sized Wireless MEMS Flow Sensor System Attachable to Medical Mask for Real-Time Respiration Monitoring — Miyoko Matsushima, Tsuyoshi Tsukada, et al. · (2025) | TGRS Research Map | TGRS