A rapid method for detecting trace Rb and Cs in high-salinity brines by ICP-MS equipped with an all-matrix sampling device

This study presents a rapid and highly sensitive method for the determination of trace rubidium (Rb) and cesium (Cs) in high-salinity brines using inductively coupled plasma mass spectrometry (ICP-MS) equipped with an all-matrix sampling (AMS) device. The AMS system achieves online gas dilution by vertically introducing argon gas into the brine sample flow, effectively reducing the severe matrix suppression effect caused by 35 g·L<sup>-1</sup> salinity to an intermediate level. Experimental results demonstrated that the signal suppression induced by coexisting cations (K<sup>+</sup>, Na<sup>+</sup>, Ca<sup>2+</sup>, Mg<sup>2+</sup>) in actual brine samples was minimal (< 1.5%), thereby eliminating the need for conventional matrix matching or standard addition approaches. Accurate quantification was achieved through a straightforward calibration process based on standard curves (Rb: 5-400 μg·L<sup>-1</sup>; Cs: 5-400 μg·L<sup>-1</sup>; R<sup>2</sup> > 0.999), enhanced by dynamic internal standard correction using yttrium (Y) and rhodium (Rh), along with optimized instrument parameters (RF power and nebulizer gas flow rate). The method demonstrates excellent limits of detection (LOD: 0.039 μg·L<sup>-1</sup> for Rb; 0.005 μg·L<sup>-1</sup> for Cs), high precision (RSD < 5%), and acceptable recovery rates (85%-108%). The accuracy was further validated through comparison with AAS standard addition for high-concentration samples (> 200 μg·L<sup>-1</sup>), yielding consistent recoveries (98.6% -114%) and inter-method deviations ≤ 12.2%. Additionally, the simplification of the sample pretreatment procedure-from a traditional multi-step dilution to a single-step dilution without acid-washed containers-enhances analytical efficiency by over 70%. This approach provides a robust, sensitive, and operationally efficient solution for the analysis of extreme high-salinity environmental samples.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-08-19
DOI
https://doi.org/10.1038/s41598-025-14552-5
Citations
1
Primary Topic
Analytical chemistry methods development
Type
article
Field-Weighted Citation Impact
0.75

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A rapid method for detecting trace Rb and Cs in high-salinity brines by ICP-MS equipped with an all-matrix sampling device

Guozhong Zhu, Xiuzhen Ma, Qi Wei, Dan Ma et al.
1 citations
Scientific Reports
Analytical chemistry methods development
0.75
article

A rapid method for detecting trace Rb and Cs in high-salinity brines by ICP-MS equipped with an all-matrix sampling device

Guozhong Zhu, Xiuzhen Ma, Qi Wei, Dan Ma, Yubin Li, Xin Liu, Zhe Ma
article en
1 citations

Abstract

This study presents a rapid and highly sensitive method for the determination of trace rubidium (Rb) and cesium (Cs) in high-salinity brines using inductively coupled plasma mass spectrometry (ICP-MS) equipped with an all-matrix sampling (AMS) device. The AMS system achieves online gas dilution by vertically introducing argon gas into the brine sample flow, effectively reducing the severe matrix suppression effect caused by 35 g·L<sup>-1</sup> salinity to an intermediate level. Experimental results demonstrated that the signal suppression induced by coexisting cations (K<sup>+</sup>, Na<sup>+</sup>, Ca<sup>2+</sup>, Mg<sup>2+</sup>) in actual brine samples was minimal (< 1.5%), thereby eliminating the need for conventional matrix matching or standard addition approaches. Accurate quantification was achieved through a straightforward calibration process based on standard curves (Rb: 5-400 μg·L<sup>-1</sup>; Cs: 5-400 μg·L<sup>-1</sup>; R<sup>2</sup> > 0.999), enhanced by dynamic internal standard correction using yttrium (Y) and rhodium (Rh), along with optimized instrument parameters (RF power and nebulizer gas flow rate). The method demonstrates excellent limits of detection (LOD: 0.039 μg·L<sup>-1</sup> for Rb; 0.005 μg·L<sup>-1</sup> for Cs), high precision (RSD < 5%), and acceptable recovery rates (85%-108%). The accuracy was further validated through comparison with AAS standard addition for high-concentration samples (> 200 μg·L<sup>-1</sup>), yielding consistent recoveries (98.6% -114%) and inter-method deviations ≤ 12.2%. Additionally, the simplification of the sample pretreatment procedure-from a traditional multi-step dilution to a single-step dilution without acid-washed containers-enhances analytical efficiency by over 70%. This approach provides a robust, sensitive, and operationally efficient solution for the analysis of extreme high-salinity environmental samples.

Scientific ReportsVol. 15(1)
China Geological Survey (CN), Xinjiang Institute of Ecology and Geography (CN), University of Chinese Academy of Sciences (CN)
Chinese Academy of Sciences
Openalex Percentile: Top 28%
Analytical chemistry methods development
0.75
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.