Development and demonstration of an active optical axis alignment method for divertor infrared thermography system in ITER

An active alignment method to compensate for optical axis misalignment between the in-vessel and ex-vessel optics caused by thermal expansion of the vacuum vessel during plasma operation was developed in this study. By observing the return light from an alignment pattern at the vacuum window and a calibration light source positioned at the first pupil plane inside the vacuum vessel, subsequent optical axis alignment during plasma operation can be conducted using only the return light from the alignment pattern. Testing with a prototype demonstrated that the developed alignment method can align the optical axis remotely and actively with accuracies of less than 0.03° for angular decentration, less than 1 mm for parallel decentration, and less than 5 mm for optical path length displacement.

Authors

Institutions

Publication Details

Journal
Review of Scientific Instruments
Published
2025-08-01
DOI
https://doi.org/10.1063/5.0273766
Primary Topic
Fusion materials and technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and demonstration of an active optical axis alignment method for divertor infrared thermography system in ITER

Ryota Imazawa, Tatsuo Sugie, Tomohiko Ushiki, Butch Buenavidez et al.
Review of Scientific Instruments
Fusion materials and technologies
article

Development and demonstration of an active optical axis alignment method for divertor infrared thermography system in ITER

Ryota Imazawa, Tatsuo Sugie, Tomohiko Ushiki, Butch Buenavidez, Naoto Kasano, Yuzi Katayanagi, Hiroyuki Tachibana, Kosuke Shimizu, Y. Nunoya, Shunsuke Tainaka
article en

Abstract

An active alignment method to compensate for optical axis misalignment between the in-vessel and ex-vessel optics caused by thermal expansion of the vacuum vessel during plasma operation was developed in this study. By observing the return light from an alignment pattern at the vacuum window and a calibration light source positioned at the first pupil plane inside the vacuum vessel, subsequent optical axis alignment during plasma operation can be conducted using only the return light from the alignment pattern. Testing with a prototype demonstrated that the developed alignment method can align the optical axis remotely and actively with accuracies of less than 0.03° for angular decentration, less than 1 mm for parallel decentration, and less than 5 mm for optical path length displacement.

Review of Scientific InstrumentsVol. 96(8)
Kai Research (United States) (US), National Institutes for Quantum Science and Technology (JP)
Openalex Percentile: Top 84%
Fusion materials and technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.