Implementation of a brief training programme to reduce mask leak with a respiratory function monitor

Respiratory function monitors (RFMs) can provide real-time data including expiratory tidal volumes and interface leak during delivery room (DR) neonatal resuscitation. However, concerns have been raised within previous randomised controlled trials and systematic reviews that effectuating RFMs routinely for DR resuscitation is limited by lack of comprehensive staff training. The International Liaison Committee on Resuscitation have also recently declared a knowledge gap in terms of the training required to be competent in utilising an RFM during neonatal resuscitation. We conducted an experiment with thirteen medical students who had no prior resuscitation experience to see if visual feedback from an RFM could improve mask leak when ventilating an extremely preterm manikin. Following a brief training session with hands-on practice, 260 breaths were analysed and median mask leak improved from 6% (IQR 1-6%) during the first breath to 0% (IQR 0-1%) by the twentieth and final breath. The median expiratory tidal volume was 3.6 mL (IQR 3.4-3.8 mL) and remained stable. This manikin experiment has demonstrated that even a short and experiential training session with an RFM for participants with no prior resuscitation training can be effective in reducing mask leak.

Authors

Institutions

Publication Details

Journal
BMJ Paediatrics Open
Published
2025-08-01
DOI
https://doi.org/10.1136/bmjpo-2025-003589
Primary Topic
Safe Handling of Antineoplastic Drugs
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Implementation of a brief training programme to reduce mask leak with a respiratory function monitor

Neaha Patel, Ludovic J. Chassin, Burak Salgin
BMJ Paediatrics Open
Safe Handling of Antineoplastic Drugs
article

Implementation of a brief training programme to reduce mask leak with a respiratory function monitor

Neaha Patel, Ludovic J. Chassin, Burak Salgin
article en

Abstract

Respiratory function monitors (RFMs) can provide real-time data including expiratory tidal volumes and interface leak during delivery room (DR) neonatal resuscitation. However, concerns have been raised within previous randomised controlled trials and systematic reviews that effectuating RFMs routinely for DR resuscitation is limited by lack of comprehensive staff training. The International Liaison Committee on Resuscitation have also recently declared a knowledge gap in terms of the training required to be competent in utilising an RFM during neonatal resuscitation. We conducted an experiment with thirteen medical students who had no prior resuscitation experience to see if visual feedback from an RFM could improve mask leak when ventilating an extremely preterm manikin. Following a brief training session with hands-on practice, 260 breaths were analysed and median mask leak improved from 6% (IQR 1-6%) during the first breath to 0% (IQR 0-1%) by the twentieth and final breath. The median expiratory tidal volume was 3.6 mL (IQR 3.4-3.8 mL) and remained stable. This manikin experiment has demonstrated that even a short and experiential training session with an RFM for participants with no prior resuscitation training can be effective in reducing mask leak.

BMJ Paediatrics OpenVol. 9(1)
Queen Mary University of London (GB), Barts Health NHS Trust (GB), William Harvey Research Institute (GB)
Nanjing Medical University
Openalex Percentile: Top 69%
Safe Handling of Antineoplastic Drugs
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.