Analysis of the Effect of Red LED and Infrared Flip Flop Frequency on SpO2 Measurement Accuracy

Oxygen saturation is a vital parameter for the early detection of advanced oxygen deficiency. Spo2 is a tool that measures the amount of oxygen in the blood non-invasively. This equipment consists of ophotodiodeiode as a sensor as well as red and infrared LEDs with a flip flop driver circuit that has a certain frequency. In this case, several research projects and equipment on the market have various flip flop frequencies. This research aims to find the best frequency setting value for red and infrared led drivers on SpO2 devices. In this research, a SpO2 that can be adjusted with a flip flop frequency of 400 Hz to 1400 Hz was designed. The SPO2 reading from the sensor is presented on the OLED LCD panel using Arduino Mega as a data processor from the driver frequency output controller. Frequency adjustment for sensor drivers is also at 400 Hz to 1400 Hz. This tool was further used to measure the frequency variation of the flip flop. The measurement results on the subject's finger were then compared with the results of the standard SpO2 tool to see the effect of the frequency value on the level of accuracy of the tool. The results of the comparison data processing showed that the largest error of 0.35% occurred in the SPO2 measurement using the 600 Hz sensor frequency driver, and the smallest error value of 0.07%, occurred in the use of the driver frequency at 1400Hz. These results can be used in the initial design of the production of SpO2 equipment, the higher the frequency, the more accurate it will be. This study only discusses the frequency, whereas the intensity parameters of the red and infrared LEDs also vary.

Authors

Institutions

Publication Details

Journal
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Published
2025-08-17
DOI
https://doi.org/10.35882/ijeeemi.v4i2.145
Primary Topic
Water Quality Monitoring and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analysis of the Effect of Red LED and Infrared Flip Flop Frequency on SpO2 Measurement Accuracy

Dyah Titisari, Moch.Prastawa Assalim T.P, Wahyu Caesarendra, Bagas Angger Prakoso
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Water Quality Monitoring and Analysis
article

Analysis of the Effect of Red LED and Infrared Flip Flop Frequency on SpO2 Measurement Accuracy

Dyah Titisari, Moch.Prastawa Assalim T.P, Wahyu Caesarendra, Bagas Angger Prakoso
article en

Abstract

Oxygen saturation is a vital parameter for the early detection of advanced oxygen deficiency. Spo2 is a tool that measures the amount of oxygen in the blood non-invasively. This equipment consists of ophotodiodeiode as a sensor as well as red and infrared LEDs with a flip flop driver circuit that has a certain frequency. In this case, several research projects and equipment on the market have various flip flop frequencies. This research aims to find the best frequency setting value for red and infrared led drivers on SpO2 devices. In this research, a SpO2 that can be adjusted with a flip flop frequency of 400 Hz to 1400 Hz was designed. The SPO2 reading from the sensor is presented on the OLED LCD panel using Arduino Mega as a data processor from the driver frequency output controller. Frequency adjustment for sensor drivers is also at 400 Hz to 1400 Hz. This tool was further used to measure the frequency variation of the flip flop. The measurement results on the subject's finger were then compared with the results of the standard SpO2 tool to see the effect of the frequency value on the level of accuracy of the tool. The results of the comparison data processing showed that the largest error of 0.35% occurred in the SPO2 measurement using the 600 Hz sensor frequency driver, and the smallest error value of 0.07%, occurred in the use of the driver frequency at 1400Hz. These results can be used in the initial design of the production of SpO2 equipment, the higher the frequency, the more accurate it will be. This study only discusses the frequency, whereas the intensity parameters of the red and infrared LEDs also vary.

Indonesian Journal of Electronics Electromedical Engineering and Medical InformaticsVol. 4(2)
Universiti Brunei Darussalam (BN), Poltekkes Kemenkes Surabaya
Openalex Percentile: Top 74%
Water Quality Monitoring and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Analysis of the Effect of Red LED and Infrared Flip Flop Frequency on SpO2 Measurement Accuracy — Dyah Titisari, Moch.Prastawa Assalim T.P, et al. · Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics (2025) | TGRS Research Map | TGRS