Accuracy of Infrared Photodiode Sensors at The Flowrate Measurement in Infusion Device Analyzer with 2 Channel TFT Display

Infusion is a medical procedure involving the delivery of fluids, medications, nutrients, or other substances into a patient's circulatory system through intravenous flow. In its implementation, commonly used infusion tools are infusion pumps and syringe pumps, as these devices assist in accurately and precisely controlling the flow rate. However, with continuous usage, there is a potential for inaccuracies in measuring the flow rate of fluids from these devices. Therefore, periodic calibration is necessary. According to Regulation No. 54 of 2015 from the Ministry of Health, calibration of medical devices, including infusion devices, must be conducted at least once a year to ensure their ongoing accuracy. The purpose of this research is to design an Infusion Device Analyzer (IDA) with a flow rate parameter. The contribution of this research is that the tool can accurately calculate the correct value of the flow rate that comes out of the infusion pump and syringe pump. One of the innovations resulting from this study is the Infusion Device Analyzer, equipped with a 7-inch TFT LCD screen that displays graphical parameters of the flow rate. This is achieved through the use of Infrared Photodiode sensors, which measure the rate of fluid flow. The performance graph displayed on the TFT LCD can visualize the stability of fluid flow when using various types of devices like syringe pumps and infusion pumps. The results of this research show a range of error values in the performance of various brands of infusion devices. When using the Terumo Syringe Pump, there is an error value of 0.86% (100 mL/h) for Channel 1 and 0.69% (100 mL/h) for Channel 2. For the B-Braun Syringe Pump, the error value is 1.30% (100 mL/h) for Channel 1 and 0.85% (10 mL/h) for Channel 2.

Authors

Institutions

Publication Details

Journal
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Published
2025-08-17
DOI
https://doi.org/10.35882/ijeeemi.v5i3.177
Primary Topic
Advanced Sensor Technologies Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy of Infrared Photodiode Sensors at The Flowrate Measurement in Infusion Device Analyzer with 2 Channel TFT Display

Triana Rahmawati, Syaifudin Syaifudin, Wafiq Nur Azizah
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Advanced Sensor Technologies Research
article

Accuracy of Infrared Photodiode Sensors at The Flowrate Measurement in Infusion Device Analyzer with 2 Channel TFT Display

Triana Rahmawati, Syaifudin Syaifudin, Wafiq Nur Azizah
article en

Abstract

Infusion is a medical procedure involving the delivery of fluids, medications, nutrients, or other substances into a patient's circulatory system through intravenous flow. In its implementation, commonly used infusion tools are infusion pumps and syringe pumps, as these devices assist in accurately and precisely controlling the flow rate. However, with continuous usage, there is a potential for inaccuracies in measuring the flow rate of fluids from these devices. Therefore, periodic calibration is necessary. According to Regulation No. 54 of 2015 from the Ministry of Health, calibration of medical devices, including infusion devices, must be conducted at least once a year to ensure their ongoing accuracy. The purpose of this research is to design an Infusion Device Analyzer (IDA) with a flow rate parameter. The contribution of this research is that the tool can accurately calculate the correct value of the flow rate that comes out of the infusion pump and syringe pump. One of the innovations resulting from this study is the Infusion Device Analyzer, equipped with a 7-inch TFT LCD screen that displays graphical parameters of the flow rate. This is achieved through the use of Infrared Photodiode sensors, which measure the rate of fluid flow. The performance graph displayed on the TFT LCD can visualize the stability of fluid flow when using various types of devices like syringe pumps and infusion pumps. The results of this research show a range of error values in the performance of various brands of infusion devices. When using the Terumo Syringe Pump, there is an error value of 0.86% (100 mL/h) for Channel 1 and 0.69% (100 mL/h) for Channel 2. For the B-Braun Syringe Pump, the error value is 1.30% (100 mL/h) for Channel 1 and 0.85% (10 mL/h) for Channel 2.

Indonesian Journal of Electronics Electromedical Engineering and Medical InformaticsVol. 5(3)
Poltekkes Kemenkes Surabaya
Openalex Percentile: Top 80%
Advanced Sensor Technologies Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Accuracy of Infrared Photodiode Sensors at The Flowrate Measurement in Infusion Device Analyzer with 2 Channel TFT Display — Triana Rahmawati, Syaifudin Syaifudin, et al. · Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics (2025) | TGRS Research Map | TGRS