Design and Development of Flow Analyzer for Peak Inspiratory Flow (PIF) and Peak Expiratory Flow (PEF) Parameters

This research study focused on the development and testing of a flow analyzer for monitoring the PIF (Peak Inspiratory Flow) and PEF (Peak Expiratory Flow) parameters in patients undergoing mechanical ventilation. These parameters are crucial for managing respiratory diseases like asthma and COPD. The study aims to ensure the accuracy of ventilator information provided to users by periodically testing and calibrating the ventilator using the flow analyzer. The primary contribution of this research is the utilization of the AFM3000 flow sensor to measure PIF and PEF parameters. By using this sensor, the researchers intend to enhance the accuracy of ventilator readings. The study was conducted in Volume Control (VC) mode with various VT (Tidal Volume) settings, ranging from 200 to 600 mL. Two different methods of data collection were employed to gather two sets of data. From the first data collection, the largest discrepancies in reading PIF and PEF values were found to be 3.49% and 2.99%, respectively. In the second data collection, the sensor exhibited a consistent reading for a constant flow of ±0.1 LPM (Liters Per Minute), indicating stability. The research findings suggest that the AFM3000 flow sensor demonstrates good accuracy and stability in measuring PIF and PEF parameters. Furthermore, the sensor is sensitive and has a minimal delay, making it suitable for real-time graph display in the module. Overall, this study contributes to the field of medical technology by developing and validating a flow analyzer for monitoring PIF and PEF parameters in mechanical ventilation. The research showcases the potential of the AFM3000 flow sensor to improve the accuracy and reliability of ventilator information, ultimately benefiting patients with respiratory disorders.

Authors

Institutions

Publication Details

Journal
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Published
2025-08-17
DOI
https://doi.org/10.35882/ijeeemi.v5i3.176
Primary Topic
Air Quality Monitoring and Forecasting
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design and Development of Flow Analyzer for Peak Inspiratory Flow (PIF) and Peak Expiratory Flow (PEF) Parameters

Levana Forra Wakidi, Endro Yulianto, Andri Lazuardi Wahyu Pambudi
5 links
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Air Quality Monitoring and Forecasting
article

Design and Development of Flow Analyzer for Peak Inspiratory Flow (PIF) and Peak Expiratory Flow (PEF) Parameters

Levana Forra Wakidi, Endro Yulianto, Andri Lazuardi Wahyu Pambudi
article en

Abstract

This research study focused on the development and testing of a flow analyzer for monitoring the PIF (Peak Inspiratory Flow) and PEF (Peak Expiratory Flow) parameters in patients undergoing mechanical ventilation. These parameters are crucial for managing respiratory diseases like asthma and COPD. The study aims to ensure the accuracy of ventilator information provided to users by periodically testing and calibrating the ventilator using the flow analyzer. The primary contribution of this research is the utilization of the AFM3000 flow sensor to measure PIF and PEF parameters. By using this sensor, the researchers intend to enhance the accuracy of ventilator readings. The study was conducted in Volume Control (VC) mode with various VT (Tidal Volume) settings, ranging from 200 to 600 mL. Two different methods of data collection were employed to gather two sets of data. From the first data collection, the largest discrepancies in reading PIF and PEF values were found to be 3.49% and 2.99%, respectively. In the second data collection, the sensor exhibited a consistent reading for a constant flow of ±0.1 LPM (Liters Per Minute), indicating stability. The research findings suggest that the AFM3000 flow sensor demonstrates good accuracy and stability in measuring PIF and PEF parameters. Furthermore, the sensor is sensitive and has a minimal delay, making it suitable for real-time graph display in the module. Overall, this study contributes to the field of medical technology by developing and validating a flow analyzer for monitoring PIF and PEF parameters in mechanical ventilation. The research showcases the potential of the AFM3000 flow sensor to improve the accuracy and reliability of ventilator information, ultimately benefiting patients with respiratory disorders.

Indonesian Journal of Electronics Electromedical Engineering and Medical InformaticsVol. 5(3)
Universitas Negeri Surabaya (ID)
Openalex Percentile: Top 81%
Air Quality Monitoring and Forecasting
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.