Experimental Measurement and Analysis: Collimation and Illumination for Conformity Measuring Instrument Design in X-ray Modality

Manual illumination and collimation testing can be affected by subjectivity. Human interpretation and judgment in measuring and adjusting illumination and collimation can vary between individuals, potentially resulting in inconsistent results. The aim of this research is to develop the simplest method for measuring illumination at four points simultaneously and directly storing the measurement data. This objective aims to address the subjectivity issues and improve the reliability and consistency of the testing process, which measures illumination at four points simultaneously and stores the measurement data directly. The method of this study was an experimental measurement and analysis that involved capturing illumination and collimation data using a suitable measuring instrument in an X-ray environment. The collected data is then analyzed to evaluate the suitability of the instrument to the established compliance standards. The module is designed using HC-SR04 sensor as a distance meter and TSL2561 sensor as a light meter. This module is designed using HC-SR04 sensor as a distance meter and TSL2561 sensor as a light meter. In this research, the module has been tested and compared with the results of the comparison tool (Digital Light Meter) and obtained an error value of 1.55% with a module efficiency of 98.45% in the illumination test, and an error of 1.8% with a module efficiency of 98.2% in the collimator test. From this research, it can be concluded that the TSL2561 light sensor can be used to measure the illumination area of the collimator lamp. The contribution of this research is expected to be as follows consistent results from tool testing, provides accuracy of results, is more efficient in cost and energy, and the data will be stored until the next testing time.

Authors

Institutions

Publication Details

Journal
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Published
2025-08-17
DOI
https://doi.org/10.35882/ijeeemi.v5i2.172
Primary Topic
Nuclear Physics and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Experimental Measurement and Analysis: Collimation and Illumination for Conformity Measuring Instrument Design in X-ray Modality

Achmad Rizal, Andjar Pudji, Muhammad Ridha Mak’ruf, Fathul Fajar
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Nuclear Physics and Applications
article

Experimental Measurement and Analysis: Collimation and Illumination for Conformity Measuring Instrument Design in X-ray Modality

Achmad Rizal, Andjar Pudji, Muhammad Ridha Mak’ruf, Fathul Fajar
article en

Abstract

Manual illumination and collimation testing can be affected by subjectivity. Human interpretation and judgment in measuring and adjusting illumination and collimation can vary between individuals, potentially resulting in inconsistent results. The aim of this research is to develop the simplest method for measuring illumination at four points simultaneously and directly storing the measurement data. This objective aims to address the subjectivity issues and improve the reliability and consistency of the testing process, which measures illumination at four points simultaneously and stores the measurement data directly. The method of this study was an experimental measurement and analysis that involved capturing illumination and collimation data using a suitable measuring instrument in an X-ray environment. The collected data is then analyzed to evaluate the suitability of the instrument to the established compliance standards. The module is designed using HC-SR04 sensor as a distance meter and TSL2561 sensor as a light meter. This module is designed using HC-SR04 sensor as a distance meter and TSL2561 sensor as a light meter. In this research, the module has been tested and compared with the results of the comparison tool (Digital Light Meter) and obtained an error value of 1.55% with a module efficiency of 98.45% in the illumination test, and an error of 1.8% with a module efficiency of 98.2% in the collimator test. From this research, it can be concluded that the TSL2561 light sensor can be used to measure the illumination area of the collimator lamp. The contribution of this research is expected to be as follows consistent results from tool testing, provides accuracy of results, is more efficient in cost and energy, and the data will be stored until the next testing time.

Indonesian Journal of Electronics Electromedical Engineering and Medical InformaticsVol. 5(2)
Politeknik Kesehatan Kemenkes Semarang (ID), Poltekkes Kemenkes Surabaya, Telkom University (ID)
Openalex Percentile: Top 72%
Nuclear Physics and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.