Design of Carbon Dioxide Levels Measurement in Human Expiration Using End Tidal Carbon Dioxide (EtCO2) Capnography Method

Asthma is a chronic respiratory disease that has become the main reason patients are always rushed to the hospital emergency department. Capnography is a new method for examining asthma by measuring CO2 levels released by the lungs. The aim of this research was to create an EtCO2 capnography device that is able to measure CO2 levels in patients with asthma or difficult breathing to assist doctors in determining the urgency of using a ventilator in a patient. The EtCO2 Capnography device used in the hospital uses a sensor that is expensive, but in this study, a CO2 gas sensor type Cozir-WX-20 was used at a low price. The research was conducted by utilizing a CO2 gas sensor type Cozir-WX-20 which read CO2 concentration in ppm value and a microcontroller as an analog to digital data processor to be displayed on the LCD. Sensor characterization was carried out to compare the side-stream and main-stream methods, response time readings, and the accuracy of the cozir sensor. The resulting data were taken from CO2 cylinders and medical air gas at various flow volume values and was connected to the Cozir sensor, EtCO2 main-stream patient monitors, and side-stream EtCO2 patient monitors. The resulting CO2 readings from CO2 tubes and medical water on the Cozir-WX-20 sensor and main-stream patient monitors obtained an error of 4.6%, namely at a CO2 concentration of 7% or 70,000 ppm and sensor accuracy is above 95%. As for the side-stream method, the reading error is 1.96% and 1.74% at a CO2 concentration of 6-7%. Sensor accuracy on the side-stream method cozir module is above 95%. Response time reading CO2 gas at a concentration of 1%-7% under 5 seconds.

Authors

Institutions

Publication Details

Journal
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Published
2025-08-17
DOI
https://doi.org/10.35882/ijeeemi.v5i1.160
Citations
1
Primary Topic
Air Quality Monitoring and Forecasting
Type
article
Field-Weighted Citation Impact
0.80
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design of Carbon Dioxide Levels Measurement in Human Expiration Using End Tidal Carbon Dioxide (EtCO2) Capnography Method

Rifky Maulana Fuadi, Endro Yulianto, Bambang Guruh Irianto, Abhishek Mishra
1 citations
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Air Quality Monitoring and Forecasting
0.80
article

Design of Carbon Dioxide Levels Measurement in Human Expiration Using End Tidal Carbon Dioxide (EtCO2) Capnography Method

Rifky Maulana Fuadi, Endro Yulianto, Bambang Guruh Irianto, Abhishek Mishra
article en
1 citations

Abstract

Asthma is a chronic respiratory disease that has become the main reason patients are always rushed to the hospital emergency department. Capnography is a new method for examining asthma by measuring CO2 levels released by the lungs. The aim of this research was to create an EtCO2 capnography device that is able to measure CO2 levels in patients with asthma or difficult breathing to assist doctors in determining the urgency of using a ventilator in a patient. The EtCO2 Capnography device used in the hospital uses a sensor that is expensive, but in this study, a CO2 gas sensor type Cozir-WX-20 was used at a low price. The research was conducted by utilizing a CO2 gas sensor type Cozir-WX-20 which read CO2 concentration in ppm value and a microcontroller as an analog to digital data processor to be displayed on the LCD. Sensor characterization was carried out to compare the side-stream and main-stream methods, response time readings, and the accuracy of the cozir sensor. The resulting data were taken from CO2 cylinders and medical air gas at various flow volume values and was connected to the Cozir sensor, EtCO2 main-stream patient monitors, and side-stream EtCO2 patient monitors. The resulting CO2 readings from CO2 tubes and medical water on the Cozir-WX-20 sensor and main-stream patient monitors obtained an error of 4.6%, namely at a CO2 concentration of 7% or 70,000 ppm and sensor accuracy is above 95%. As for the side-stream method, the reading error is 1.96% and 1.74% at a CO2 concentration of 6-7%. Sensor accuracy on the side-stream method cozir module is above 95%. Response time reading CO2 gas at a concentration of 1%-7% under 5 seconds.

Indonesian Journal of Electronics Electromedical Engineering and Medical InformaticsVol. 5(1)
Rama University (IN), Poltekkes Kemenkes Surabaya
Openalex Percentile: Top 26%
Air Quality Monitoring and Forecasting
0.80
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.