Pressure Sensor Stability Analysis of Positive End Expiratory Pressure Parameters in Flow Analyzer Design

The Positive End Expiratory Pressure (PEEP) parameter is a parameter that must be considered in the process of determining the patient's condition, a safe threshold, and must be in accordance with the settings. However, the PEEP value on the ventilator often does not match the settings so that the measuring instrument capable of detecting PEEP on the ventilator is the Flow Analyzer. The purpose of this study was to design a Flow Analyzer using the MPX2010 sensor to analyze the stability of the PEEP parameters on the ventilator. The main contribution of this research is the design of a simple Flow Analyzer device with stable monitoring of PEEP parameters and the availability of many required setting options. This study used PEEP settings of 0, 5, 8, 11, 14, 17, 20, 23, 26, and 29 cmH2O. In this case, data were collected using a ventilator with VCV (Volume Control Ventilation) and PCV (Pressure Control Ventilation) modes. The tool used for reference from standard measurements was the Standard Flow Analyzer tool. The results of this study indicated that the measurement accuracy of PEEP parameters with the Flow Analyzer module at each PEEP setting had the smallest error of ±0% at 0 cmH2O setting so that it also had the smallest value of 0 by standard. deviation and uncertainty (UA) value 0 at each setting. Meanwhile, the Flow Analyzer measurement module had the largest error in the 5 cmH2O setting, which was ±13.2% with the largest correction value of 0.77. Based on the data obtained, the monitoring of the PEEP parameter was considered quite stable even though the value was still out of tolerance. Therefore, the monitoring of PEEP stability parameters can be implemented during the ventilator calibration process in order to analyze damage and reduce the time of damage to the ventilator.

Authors

Institutions

Publication Details

Journal
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Published
2025-08-17
DOI
https://doi.org/10.35882/ijeeemi.v5i1.164
Primary Topic
Air Quality Monitoring and Forecasting
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Pressure Sensor Stability Analysis of Positive End Expiratory Pressure Parameters in Flow Analyzer Design

Eko Dedi Prasetyo, P. Chandrasekaran, Levana Forra Wakidi, Bambang Guruh Irianto et al.
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Air Quality Monitoring and Forecasting
article

Pressure Sensor Stability Analysis of Positive End Expiratory Pressure Parameters in Flow Analyzer Design

Eko Dedi Prasetyo, P. Chandrasekaran, Levana Forra Wakidi, Bambang Guruh Irianto, Abd Kholiq
article en

Abstract

The Positive End Expiratory Pressure (PEEP) parameter is a parameter that must be considered in the process of determining the patient's condition, a safe threshold, and must be in accordance with the settings. However, the PEEP value on the ventilator often does not match the settings so that the measuring instrument capable of detecting PEEP on the ventilator is the Flow Analyzer. The purpose of this study was to design a Flow Analyzer using the MPX2010 sensor to analyze the stability of the PEEP parameters on the ventilator. The main contribution of this research is the design of a simple Flow Analyzer device with stable monitoring of PEEP parameters and the availability of many required setting options. This study used PEEP settings of 0, 5, 8, 11, 14, 17, 20, 23, 26, and 29 cmH2O. In this case, data were collected using a ventilator with VCV (Volume Control Ventilation) and PCV (Pressure Control Ventilation) modes. The tool used for reference from standard measurements was the Standard Flow Analyzer tool. The results of this study indicated that the measurement accuracy of PEEP parameters with the Flow Analyzer module at each PEEP setting had the smallest error of ±0% at 0 cmH2O setting so that it also had the smallest value of 0 by standard. deviation and uncertainty (UA) value 0 at each setting. Meanwhile, the Flow Analyzer measurement module had the largest error in the 5 cmH2O setting, which was ±13.2% with the largest correction value of 0.77. Based on the data obtained, the monitoring of the PEEP parameter was considered quite stable even though the value was still out of tolerance. Therefore, the monitoring of PEEP stability parameters can be implemented during the ventilator calibration process in order to analyze damage and reduce the time of damage to the ventilator.

Indonesian Journal of Electronics Electromedical Engineering and Medical InformaticsVol. 5(1)
Universitas Negeri Surabaya (ID)
Openalex Percentile: Top 81%
Air Quality Monitoring and Forecasting
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.